In plain language
In 80 fenestrated aortic repairs by one surgeon, real-time dosimetry showed lower doses to the surgeon, trainee and nurses with standard fenestrated and off-the-shelf devices than with custom-made devices, and higher operator dose with four-vessel repairs.
At a glance
- Type of study
- Original research, Prospective cohort study
- Who was studied
- Vascular surgeons, Trainees, Anesthesia staff, Nurses, Patients
- Size
- n = 80 procedures
- Summary based on
- the abstract
What it found
Zenith Fenestrated devices had lower patient and staff doses than custom-made devices.
Key figures: operator 120 μSv vs 370 μSv; P = .004; patient 1800 mGy vs 2950 mGy; P = .004
Off-the-shelf branched devices had lower operator, assistant and circulator doses than custom-made devices.
Key figures: operator 220 μSv vs 370 μSv; P = .04; circulator 4 μSv vs 30 μSv; P = .001
Four-vessel fenestrated repairs gave higher operator dose than three-vessel repairs.
Key figures: 440 μSv vs 170 μSv; P = .0005
Nurse doses were much lower than operator and assistant doses.
Key figures: circulator 10 μSv vs 30 μSv; scrub 10 μSv vs 40 μSv
Authors judged all doses within acceptable limits and lower than previously reported.
Limits to keep in mind
Single-center, single-surgeon series on one fluoroscopy system with dose-reduction technology, with small numbers in the ZFen and OTS groups and no adjustment beyond body mass index (BMI); device groups were not randomized.
Related research summaries
- Comparison of Radiation Exposure during Endovascular Treatment of Peripheral Arterial Disease with Flat-Panel Detectors on Mobile C-arm versus Fixed Systems (2017, Prospective cohort study)
- ICRP Draft - Radiological Protection in Fluoroscopically Guided Procedures (Guideline or consensus statement)
- Radiation exposure to operating room personnel and patients during endovascular procedures (2013, Dosimetry measurement study)
- AAPM Medical Physics Practice Guideline 12.a - Fluoroscopy Dose Management (2022, Guideline or consensus statement)
- Modern Fixed Imaging Systems Reduce Radiation Exposure To Patients And Providers (2017, Retrospective cohort study)
How this summary was made
ORSIF prepared this summary with AI assistance from the abstract of the paper. A program then checked that the passage supporting each finding appears word for word in that source. A person has not yet reviewed each summary, so read the original before relying on a number. ORSIF summarizes research and does not endorse any product or company. Report a problem with this summary.
Prepared November 5, 2026.