In plain language
This US medical physics guideline recommends cumulative air kerma for patient dose alerts (first at 3 Gy, then every 1 Gy), follow-up above 5 Gy, and single collar-badge monitoring with lens dose assessment for the fluoroscopy team.
At a glance
- Type of study
- Guideline or consensus statement
- Who was studied
- Medical physicists and radiation safety officers, Patients, Interventional radiologists, Interventional cardiologists, Vascular surgeons, Radiologic technologists, Nurses, Anesthesia staff, Mixed lab staff
- Where
- United States
- Topics
- Patient doseDosimetry and monitoringRegulation and policyEye lens and cataractsTraining and educationDose reduction techniquesSafety culture and compliance
- Summary based on
- the full text
What it found
Cumulative air kerma (Ka,r) is recommended as the primary dose index for setting notification and patient follow-up thresholds.
Consensus intra-procedural notification levels start at 3 Gy Ka,r, with further notifications every 1 Gy Ka,r.
Key figures: first at 3 Gy Ka,r; every 1 Gy thereafter
Suggested SRDL for post-procedure patient follow-up is 5 Gy Ka,r or 3 Gy PSD, with doses summed over a 60-day window.
Key figures: 5 Gy Ka,r or 3 Gy PSD; 60 days
A single collar badge with personal protective equipment (PPE) correction is recommended for fluoroscopically guided interventional (FGI) staff because double badging is often misused; over a third of two-badge entries were invalid in a large review.
Key figures: over a third of entries invalid
Lens dose should be assessed; a prior large study found over 15% of workers exceeded the International Commission on Radiological Protection (ICRP) 20 mSv/year lens limit.
Key figures: over 15%; 20 mSv/year
Limits to keep in mind
Consensus practice guideline based on expert opinion and cited literature rather than new data; it focuses mainly on patient tissue reactions, with occupational monitoring covered only briefly.
Related research summaries
- Radiation Safety and Protection - StatPearls (2024, Narrative review)
- ICRP Draft - Radiological Protection in Fluoroscopically Guided Procedures (Guideline or consensus statement)
- Occupational Protection in Interventional Radiology. A Joint Guideline of the Cardiovascular and Interventional Radiological Society of Europe and the Society of Interventional Radiology (2026, Guideline or consensus statement)
- AST Standards for Ionizing Radiation Exposure (Guideline or consensus statement)
- Safety of Fluoroscopy - StatPearls (2024, Narrative review)
How this summary was made
ORSIF prepared this summary with AI assistance from the full text of the paper. A program then checked that the passage supporting each finding appears word for word in that source. A person has not yet reviewed each summary, so read the original before relying on a number. ORSIF summarizes research and does not endorse any product or company. Report a problem with this summary.
Prepared November 5, 2026.