In lower-limb endovascular procedures, a mobile flat-panel C-arm was associated with lower patient dose than a fixed system. Surgeon and nurse doses were similar, but trainee residents and nurse anesthetists recorded higher doses with the mobile C-arm.
At a glance
- Type of study
- Original research, Prospective cohort study
- Who was studied
- Vascular surgeons, Nurses, Trainees, Anesthesia staff, Patients
- Size
- n = 419 patients
- Summary based on
- the abstract
What it found
After adjustment for body mass index, patient air kerma was lower with the mobile C-arm than with the fixed C-arm.
Key figures: 124.1 ± 142 vs. 173.3 ± 248.7 mGy, P = 0.025
Effective dose to senior surgeons and operating room nurses did not differ between mobile and fixed systems.
The trainee resident and nurse anesthetists recorded a higher effective dose on the external dosimeter with the mobile C-arm.
The authors conclude the type of imaging system does not appear to affect staff exposure overall.
Limits to keep in mind
Nonrandomized groups with unequal sizes and a small staff-dose series (24 vs 76 patients); abstract only, with no staff dose values or country reported.
Related research summaries
- ICRP Draft - Radiological Protection in Fluoroscopically Guided Procedures (Guideline or consensus statement)
- Radiation exposure to operating room personnel and patients during endovascular procedures (2013, Dosimetry measurement study)
- AAPM Medical Physics Practice Guideline 12.a - Fluoroscopy Dose Management (2022, Guideline or consensus statement)
- Patient And Operating Room Staff Radiation Dose During Fenestrated/Branched Endovascular Aneurysm Repair Using Premanufactured Devices (2018, Prospective cohort study)
- Modern Fixed Imaging Systems Reduce Radiation Exposure To Patients And Providers (2017, Retrospective cohort study)
How this summary was made
ORSIF prepared this summary with AI assistance from the abstract of the paper. A program then checked that the passage supporting each finding appears word for word in that source. A person has not yet reviewed each summary, so read the original before relying on a number. ORSIF summarizes research and does not endorse any product or company. Report a problem with this summary.
Prepared November 5, 2026.