In plain language
In 657 cardiac device implantations across nine centers, cardiac resynchronization therapy (CRT) procedures carried the highest patient dose, and newer customizable X-ray systems roughly tripled-down adjusted dose. Staff dose was measured but results are not given in the abstract.
At a glance
- Type of study
- Original research, Prospective cohort study
- Who was studied
- Electrophysiologists, Mixed lab staff, Patients
- Size
- n = 657 procedures
- Summary based on
- the abstract
What it found
Biventricular (CRT) device implantation had far higher fluoroscopy time and dose-area product (DAP) than dual chamber or ventricular device implantation.
Key figures: Median FT 837 s (IQR 480-1323) vs 117 s and 101 s; median DAP 1410 (IQR 807-2601) vs 150 and 129 cGy.cm²
During CRT, newer customizable X-ray systems gave lower adjusted DAP than four older systems.
Key figures: adjusted DAP 0.90 (IQR 0.26-1.01) vs 0.29 (IQR 0.23-0.39); P < .001
Dose reference levels for device implantation were proposed as third-quartile values from real-life data.
Fluoroscopy time during biventricular implantation was shorter than previously reported values.
Limits to keep in mind
Observational registry with abstract only; staff dosimeter results are not reported in the abstract, and system comparisons may be confounded by center practices.
Related research summaries
- Reducing radiation exposure during procedures performed in the electrophysiology laboratory (2017, Retrospective cohort study)
- 2018 ACC/HRS/NASCI/SCAI/SCCT Expert Consensus Document on Optimal Use of Ionizing Radiation in Cardiovascular Imaging: Best Practices for Safety and Effectiveness: A Report of the American College of Cardiology Task Force on Expert Consensus Decision Pathways (2019, Guideline or consensus statement)
- Practical ways to reduce radiation dose for patients and staff during device implantations and electrophysiological procedures (2014, Guideline or consensus statement)
- Lowering Fluoroscopy Pulse Rates To Reduce Radiation Dose During Cardiac Procedures (2018, Editorial or commentary)
- Fluoroscopy-minimization strategy for catheter ablation of Supraventricular Tachycardia by wOmen OPerators: the STOOP Multicentre Registry (2026, Prospective cohort study)
How this summary was made
ORSIF prepared this summary with AI assistance from the abstract of the paper. A program then checked that the passage supporting each finding appears word for word in that source. A person has not yet reviewed each summary, so read the original before relying on a number. ORSIF summarizes research and does not endorse any product or company. Report a problem with this summary.
Prepared November 5, 2026.