Research summary, original research, 2018

Radiation Doses For Endovascular Aortic Repairs Performed On Mobile And Fixed C-Arm Fluoroscopes And Procedure Phase-Specific Radiation Distribution

Schaefers JF, Wunderle K, Usai MV, Torsello GF, Panuccio G. Journal of vascular surgery. 2018

Read the original paper

In plain language

In 160 endovascular aneurysm repair (EVAR) patients, fixed C-arm procedures used less contrast but had about three times the dose-area product of mobile C-arm procedures; digital subtraction angiography produced nearly half the exposure. Staff dose was not measured.

At a glance

Type of study
Original research, Retrospective cohort study
Who was studied
Patients
Size
n = 160 patients
Summary based on
the abstract

What it found

  • Fixed C-arm EVAR procedures had more than three times the dose-area product of mobile C-arm procedures.

    Key figures: 49.93 ± 38.06 Gy·cm2 (mobile) vs 168.34 ± 146.92 Gy·cm2 (fixed); P < .0001

  • Fixed C-arm use required less contrast agent.

    Key figures: 108.55 ± 42.28 mL vs 85.37 ± 38.79 mL; P = .0014

  • Fluoroscopy time did not differ significantly between groups.

  • Proximal landing zone identification and main body deployment used the most radiation.

    Key figures: 24% of total dose

  • Digital subtraction angiography accounted for nearly half of exposure.

    Key figures: 47.6%

Limits to keep in mind

Non-randomized single-centre comparison of patient dose-area product only; no staff dosimetry was measured, and phase analysis was available only from the fixed system. Country not stated in the abstract.

How this summary was made

ORSIF prepared this summary with AI assistance from the abstract of the paper. A program then checked that the passage supporting each finding appears word for word in that source. A person has not yet reviewed each summary, so read the original before relying on a number. ORSIF summarizes research and does not endorse any product or company. Report a problem with this summary.

Prepared November 3, 2026.