In plain language
In 160 endovascular aneurysm repair (EVAR) patients, fixed C-arm procedures used less contrast but had about three times the dose-area product of mobile C-arm procedures; digital subtraction angiography produced nearly half the exposure. Staff dose was not measured.
At a glance
- Type of study
- Original research, Retrospective cohort study
- Who was studied
- Patients
- Size
- n = 160 patients
- Summary based on
- the abstract
What it found
Fixed C-arm EVAR procedures had more than three times the dose-area product of mobile C-arm procedures.
Key figures: 49.93 ± 38.06 Gy·cm2 (mobile) vs 168.34 ± 146.92 Gy·cm2 (fixed); P < .0001
Fixed C-arm use required less contrast agent.
Key figures: 108.55 ± 42.28 mL vs 85.37 ± 38.79 mL; P = .0014
Fluoroscopy time did not differ significantly between groups.
Proximal landing zone identification and main body deployment used the most radiation.
Key figures: 24% of total dose
Digital subtraction angiography accounted for nearly half of exposure.
Key figures: 47.6%
Limits to keep in mind
Non-randomized single-centre comparison of patient dose-area product only; no staff dosimetry was measured, and phase analysis was available only from the fixed system. Country not stated in the abstract.
Related research summaries
- Fluoroscopy-minimization strategy for catheter ablation of Supraventricular Tachycardia by wOmen OPerators: the STOOP Multicentre Registry (2026, Prospective cohort study)
- Two Central Nervous System Tumors in One Catheter Lab: Time to Rethink Radiation Protection (2026, Retrospective cohort study)
- Radiation Awareness for EVAR in Hybrid Operating Room: An Instant Operator Risk Chart (2021, Dosimetry measurement study)
- Decreased Operator X-Ray Exposure By Optimized Fluoroscopy During Radiofrequency Ablation Of Common Atrial Flutter (2018, Prospective cohort study)
- Factors Contributing To Radiation Dose For Patients And Operators During Diagnostic Cardiac Angiography (2018, Retrospective cohort study)
How this summary was made
ORSIF prepared this summary with AI assistance from the abstract of the paper. A program then checked that the passage supporting each finding appears word for word in that source. A person has not yet reviewed each summary, so read the original before relying on a number. ORSIF summarizes research and does not endorse any product or company. Report a problem with this summary.
Prepared November 3, 2026.