In plain language
In 50 endovascular aneurysm repair (EVAR) procedures in one hybrid OR, operator dose rates rose disproportionately with C-arm rotation and roughly 2.5-3 fold with brachial versus femoral access. These are over-apron dosimeter readings, not health outcomes.
At a glance
- Type of study
- Original research, Dosimetry measurement study
- Who was studied
- Vascular surgeons, Patients
- Size
- n = 12,865 X-ray acquisitions (from 50 patients)
- Summary based on
- the full text
What it found
Switching from femoral to brachial access markedly increased operator dose rate, far more than patient dose rate.
Key figures: Odds ratio (OR) 2.5 [2.32-2.67] for operator vs 1.05 for patient (fluoroscopy)
C-arm rotation increased operator dose rate disproportionately compared with patient dose rate.
Key figures: 69% (fluoroscopy) and 85% (DSA) per 30° vs 24% and 37% for patient
Predicted time to reach 20 mSv annual limit during 2 fps DSA was much shorter from brachial than femoral access.
Key figures: 941 minutes femoral vs 358 minutes brachial
DSA acquisitions gave much higher operator dose rates than fluoroscopy.
Key figures: median 0.833 mSv/h DSA vs 0.119 mSv/h fluoroscopy
Replacing 2 fps DSA with a fluoroscopy roadmap reduced operator dose rate.
Key figures: effect size 0.45 (55% reduction)
Limits to keep in mind
Single hybrid OR, 50 patients, two operators with a single over-apron chest dosimeter (no eye lens dose, positions and shield placement not recorded); low-dose runs below detector range were excluded, and operator models explained limited variance (r2 0.38-0.48).
Related research summaries
- Surgeon Education Decreases Radiation Dose In Complex Endovascular Procedures And Improves Patient Safety (2013, Retrospective cohort study)
- ICRP Publication 117. Radiological Protection In Fluoroscopically Guided Procedures Performed Outside The Imaging Department (2010, Guideline or consensus statement)
- ICRP Draft - Radiological Protection in Fluoroscopically Guided Procedures (Guideline or consensus statement)
- AAPM Medical Physics Practice Guideline 12.a - Fluoroscopy Dose Management (2022, Guideline or consensus statement)
- IU Health Handbook of Fluoroscopy Safety (Guideline or consensus statement)
How this summary was made
ORSIF prepared this summary with AI assistance from the full text of the paper. A program then checked that the passage supporting each finding appears word for word in that source. A person has not yet reviewed each summary, so read the original before relying on a number. ORSIF summarizes research and does not endorse any product or company. Report a problem with this summary.
Prepared November 3, 2026.