Research summary, original research, 2017

Radiation dose to the operator during fluoroscopically guided spine procedures

Roccatagliata L, Presilla S, Pravatà E, Cianfoni A. Neuroradiology. 2017

Read the original paper

In plain language

In 57 fluoroscopy-guided spine procedures, operator dose had a median of 6.4 μGy per procedure and tracked patient kerma-area product (KAP) moderately well (r2 0.60). This makes KAP a rough estimate of operator exposure, not a health outcome measure.

At a glance

Type of study
Original research, Dosimetry measurement study
Who was studied
Other physicians, Patients
Size
n = 57 procedures
Summary based on
the abstract

What it found

  • Operator air kerma varied widely across spine procedures, with a median of 6.4 μGy per procedure.

    Key figures: median 6.4 μGy per procedure

  • Operator air kerma correlated with patient KAP, at about 1.6 μGy operator air kerma per Gy cm2 KAP.

    Key figures: r2 = 0.60; slope 1.6 μGy per Gy cm2

  • Median fluoroscopy time was 2.6 min and median KAP 4.7 Gy cm2 per procedure; fluoroscopy time correlated with KAP.

    Key figures: 2.6 min; 4.7 Gy cm2; r2 = 0.63

  • Operator dose can be approximated from routinely recorded patient KAP.

Limits to keep in mind

Small retrospective single-series study of 57 procedures using one electronic dosimeter; dosimeter position, shielding use, operator specialty and country are not reported in the abstract.

How this summary was made

ORSIF prepared this summary with AI assistance from the abstract of the paper. A program then checked that the passage supporting each finding appears word for word in that source. A person has not yet reviewed each summary, so read the original before relying on a number. ORSIF summarizes research and does not endorse any product or company. Report a problem with this summary.