Research summary, original research, 2013

Radiation Dose To Procedural Personnel And Patients From An X-Ray Volume Imaging System

Paul J, Mbalisike EC, Vogl TJ. European radiology. 2013

Read the original paper

In plain language

In 40 interventional procedures on one X-ray volume imaging system, the interventionist's hands and left knee received more dose than the assistant physician's. Hands received the most of the regions measured. Cone-beam computed tomography (CT) gave patients the highest dose.

At a glance

Type of study
Original research, Dosimetry measurement study
Who was studied
Interventional radiologists, Other physicians, Patients
Size
n = 40 patients
Summary based on
the abstract

What it found

  • Hand and left-knee doses to the interventionist were higher than those to the assistant physician.

    Key figures: P < 0.05

  • Of the anatomical regions measured, the hands of procedural personnel received the highest doses.

  • Patient dose (DAP and skin entry dose) was lowest with catheter angiography and highest with cone-beam CT, with significant differences between groups.

    Key figures: all P = 0.0001

  • Cone-beam computed tomography (CBCT) allowed clear lesion delineation, but objective and subjective image-quality assessments gave opposite results because of noise and artefacts.

    Key figures: all P < 0.05

Limits to keep in mind

This is a small single-system study of 40 patients, and only the abstract was available. Staff dose values, the number of personnel, the country and the shielding used are not reported.

How this summary was made

ORSIF prepared this summary with AI assistance from the abstract of the paper. A program then checked that the passage supporting each finding appears word for word in that source. A person has not yet reviewed each summary, so read the original before relying on a number. ORSIF summarizes research and does not endorse any product or company. Report a problem with this summary.