In plain language
A survey of 4219 patients in 33 French centres set patient dose reference levels for common electrophysiology and pacing procedures; complex devices and left bundle branch area pacing used more radiation. Staff dose was not measured.
At a glance
- Type of study
- Original research, Cross-sectional survey
- Who was studied
- Patients
- Size
- n = 4,219 patients
- Where
- France
- Summary based on
- the abstract
What it found
Interventional reference levels were proposed for 11 common electrophysiology and pacing procedures from 4219 patients in 33 French centres.
Key figures: 4219 patients; 33 centres; 11 procedures
For device implantation, reference levels rose with procedure complexity, from single-chamber to bi-ventricular devices.
Key figures: 4.10 Gy.cm2 and 5.5 min (single-chamber pacemaker) vs 11.50 Gy.cm2 and 16.5 min (bi-ventricular pacemaker); 12.90 Gy.cm2 and 18.0 min (bi-ventricular defibrillator)
Left bundle branch area pacing significantly increased dose indicators and fluoroscopy time across pacemaker procedures.
Key figures: p < 0.05
Among ablations, accessory pathway ablation had the lowest and persistent atrial fibrillation ablation the highest reference levels.
Key figures: 4.10 Gy.cm2 and 10.5 min vs 12.65 Gy.cm2 and 15.5 min
Proposed reference levels were similar to or lower than previously published values for most procedures.
Limits to keep in mind
Only patient dose indicators were recorded; staff dose was not measured. Data come from a sample of 10 to 30 patients per procedure per centre in France, and the abstract gives limited detail on equipment variation.
Related research summaries
- Two Central Nervous System Tumors in One Catheter Lab: Time to Rethink Radiation Protection (2026, Retrospective cohort study)
- [The Multicenter Radiation Dose Survey for Catheter Ablation of Arrhythmia: Actual Conditions and Technique-specific Comparison after the DRLs 2020 Revision] (2026, Retrospective cohort study)
- Establishing updated diagnostic reference levels for interventional radiology: a national Italian survey incorporating procedure complexity indices - Part II: interventional neuroradiology (2026, Dosimetry measurement study)
- Investigation Of Reference Levels And Radiation Dose Associated With Abdominal EVAR (Endovascular Aneurysm Repair) Procedures Across Several European Centres (2017, Retrospective cohort study)
- Patient dose in interventional radiology: a multicentre study of the most frequent procedures in France (2017, Retrospective cohort study)
How this summary was made
ORSIF prepared this summary with AI assistance from the abstract of the paper. A program then checked that the passage supporting each finding appears word for word in that source. A person has not yet reviewed each summary, so read the original before relying on a number. ORSIF summarizes research and does not endorse any product or company. Report a problem with this summary.
Prepared November 3, 2026.