Research summary, original research, 2012

Operator-Controlled Imaging Significantly Reduces Radiation Exposure During EVAR

Peach G, Sinha S, Black SA, Morgan RA, Loftus IM, Thompson MM, et al.. European journal of vascular and endovascular surgery : the official journal of the European Society for Vascular Surgery. 2012

Read the original paper

In plain language

In 122 elective endovascular aneurysm repair (EVAR) cases at one centre, letting the operating surgeon control imaging was linked to lower patient dose area product and shorter fluoroscopy time. Staff dose was not measured, though lower dose-area product (DAP) usually means less scatter for the team.

At a glance

Type of study
Original research, Retrospective cohort study
Who was studied
Patients, Vascular surgeons, Radiologic technologists
Size
n = 122 patients
Summary based on
the abstract

What it found

  • Median dose area product was lower after the switch to operator-controlled imaging than before.

    Key figures: 4.9 mGy m(2) (range 1.25-13.3) vs 6.9 mGy m(2) (range 1.91-95.0), p = 0.005

  • Median screening (fluoroscopy) time fell after operator-controlled imaging was introduced.

    Key figures: 20.0 min vs 16.2 min, p = 0.027

  • Contrast volume did not differ significantly.

    Key figures: 100 ml vs 90 ml, p = 0.21

  • The reduction was particularly seen in the number of higher-dose cases.

Limits to keep in mind

Single-centre retrospective before-after comparison of 122 patients, open to time trends and learning effects. Only patient DAP and screening time were measured, not staff dose.

How this summary was made

ORSIF prepared this summary with AI assistance from the abstract of the paper. A program then checked that the passage supporting each finding appears word for word in that source. A person has not yet reviewed each summary, so read the original before relying on a number. ORSIF summarizes research and does not endorse any product or company. Report a problem with this summary.