In plain language
In six cardiac cath lab nurses in Japan, eye lens doses reached up to about 9 mSv per year without lead glasses. Neck badges and patient dose metrics correlated with eye dose but imprecisely, so dedicated eye dosimeters are advised for higher-exposed nurses.
At a glance
- Type of study
- Original research, Dosimetry measurement study
- Who was studied
- Nurses
- Size
- n = 6 nurses
- Where
- Japan
- Summary based on
- the full text
What it found
The highest-exposed nurse, who assisted in the most procedures, had projected annual eye doses below the International Commission on Radiological Protection (ICRP) limit but within about 5 mSv of it.
Key figures: DOSIRIS left and right eye = 8.9 and 8.4 mSv/y; neck 13.2 and 15.4 mSv/y
Lens and neck badge doses correlated significantly, but neck badges tended to read higher and many values fell outside the 95% confidence interval (CI), so direct eye dosimetry is recommended.
Key figures: R2 = 0.6968 (left), 0.6508 (right)
No significant left-right difference in nurse eye or neck doses; right side was slightly higher.
Key figures: lens p = 0.8769; neck p = 0.7038
Nurse neck and lens doses correlated with patient fluoroscopy time and air kerma, more weakly for lens dose.
Key figures: neck r ≥ 0.8; lens r < 0.8
The nurse with the highest per-procedure dose may have used the protective plate improperly and stayed close to the patient.
Limits to keep in mind
Only six nurses at a single Japanese hospital over 6 months, without lead glasses; correlations based on monthly aggregate readings with variation across procedure types and nurse behavior.
Related research summaries
- Occupational radiation exposure and crystalline lens density in perioperative nurses: a short communication (2026, Cross-sectional survey)
- Radiation protection: safety measures and knowledge among interventional radiologists - UK-based analysis (2025, Cross-sectional survey)
- Investigation of Radiation Exposure of Medical Staff During Lateral Fluoroscopy for Posterior Spinal Fusion Surgery (2024, Phantom or simulation study)
- Radiation eye dose to medical staff during respiratory endoscopy under X-ray fluoroscopy (2020, Dosimetry measurement study)
- Radiation-associated lens opacities in catheterization personnel: results of a survey and direct assessments (2013, Cross-sectional survey)
How this summary was made
ORSIF prepared this summary with AI assistance from the full text of the paper. A program then checked that the passage supporting each finding appears word for word in that source. A person has not yet reviewed each summary, so read the original before relying on a number. ORSIF summarizes research and does not endorse any product or company. Report a problem with this summary.
Prepared November 5, 2026.