In a phantom study, radiation passing through 0.25 mmPb lead aprons rose linearly with defect size, and an automated video tool helped set a discard threshold of about 2.87 cm2 for slit-like defects under endoscopic retrograde cholangiopancreatography (ERCP)-like conditions.
At a glance
- Type of study
- Original research, Phantom or simulation study
- Who was studied
- Phantom or simulation (no people)
- Size
- n = 18 lead aprons
- Summary based on
- the abstract
What it found
Using a 5 mSv per 3 months dose limit and a baseline transmitted dose of 2.38 mSv, the discard threshold for slit-like apron defects was 2.87 cm2.
Key figures: 2.87 cm2
Transmitted dose through the apron increased linearly with defect area under all tested conditions in an ERCP-mimicking phantom setup.
A fine-tuned YOLOv8 segmentation model detected apron defects on fluoroscopic video with modest accuracy and processed each video in 15 seconds.
Key figures: AP@0.5 of 0.457; 15 s processing time
Slit widths of 1.0, 2.0 and 3.5 mm behaved similarly in dose terms and were pooled for modelling.
Key figures: 1.0, 2.0, and 3.5 mm
Limits to keep in mind
Phantom dosimetry with artificial slit defects under ERCP-mimicking conditions; only 18 aprons of one lead equivalence, and the detection model's precision was modest (AP@0.5 0.457). The threshold depends on assumed baseline dose and regulatory limit.
Related research summaries
- Effect of radiation protection apron defect on radiation dose during fluoroscopy (2026, Dosimetry measurement study)
- Insights into Testicular Radiation Exposure in the Orthopedic Surgeon (2026, Phantom or simulation study)
- Lightweight X-ray shielding drape with a detachable shielding wall for enhanced occupational radiation protection in interventional radiology (2026, Phantom or simulation study)
- Occupational Radiation Exposure in Spine Surgery: Organ-specific OSL Phantom Dosimetry and Workload-based Risk Assessment (2026, Phantom or simulation study)
- Ocular radiation protection efficiency and ergonomic performance of a novel Shoulder-Mounted radiation shield for interventional procedures (2026, Phantom or simulation study)
How this summary was made
ORSIF prepared this summary with AI assistance from the abstract of the paper. A program then checked that the passage supporting each finding appears word for word in that source. A person has not yet reviewed each summary, so read the original before relying on a number. ORSIF summarizes research and does not endorse any product or company. Report a problem with this summary.
Prepared November 3, 2026.