Lightweight X-ray shielding drape with a detachable shielding wall for enhanced occupational radiation protection in interventional radiology

Sagehashi K, Sasaki E, Ishida M, Imai H, Kanke K, Sato T, et al.. Radiation Physics and Chemistry. 2026

Read the original paper

In plain language

In a phantom simulating coronary intervention, a lightweight patient drape with a detachable shielding wall cut measured scatter dose by 70-80% at operator eye and hand positions, though protection varied with projection angle. Clinical staff dose was not measured.

At a glance

Type of study
Original research, Phantom or simulation study
Who was studied
Phantom or simulation (no people)
Summary based on
the abstract

What it found

  • The drape with detachable shielding wall reduced dose by 70-80% at the forward-leaning eye lens and hand positions.

    Key figures: 70-80%

  • Adding the shielding wall gave more protection than the drape alone.

  • Shielding effect depended on projection angle and was limited at the eye lens position for some left anterior oblique (LAO) and caudal projections.

  • Vertical and forward-tilted wall configurations did not differ significantly.

  • The device is lightweight with low lead equivalence.

    Key figures: 0.125 mmPb; 938 g

Limits to keep in mind

Phantom study using ambient dose equivalent during simulated percutaneous coronary intervention (PCI), not personal dosimetry on staff in real procedures; abstract only, without detailed statistics.

How this summary was made

ORSIF prepared this summary with AI assistance from the abstract of the paper. A program then checked that the passage supporting each finding appears word for word in that source. A person has not yet reviewed each summary, so read the original before relying on a number. ORSIF summarizes research and does not endorse any product or company. Report a problem with this summary.