The Evidence
Resources
Clinician AssessmentGet Your Risk Score FAQs Admin ToolkitMembers Economic Impact Research ArticlesMembers SIRP 2025 WebinarMembers Videos
Guidelines
What is ALARA+? US Guidelines International Guidelines News Is Not Here About Contact Donate
Join Free

Please double-check this information. This is an ORSIF research summary, researched November 3, 2026. It is not comprehensive and not always accurate: rules change, and answers may be incomplete, out of date or wrong. It is education, not legal advice. Not yet reviewed by a regulatory expert. Confirm each requirement at the source linked beside the answer, with the state program and with your facility's radiation safety officer. How this was made

In brief

ORSIF's research identified the Texas Department of State Health Services (DSHS), Radiation Control Program as the agency that regulates X-ray and fluoroscopy equipment in Texas. Texas is a U.S. Nuclear Regulatory Commission (NRC) Agreement State: it has signed an agreement with the NRC to regulate certain radioactive materials itself.

Texas (DSHS Radiation Control Program, Agreement State since 1963) requires ALARA for X-ray registrants in a machine-only rule (25 TAC §289.231), separate from the materials rule (§289.202), plus an operator-position ALARA line in §289.227. Dose limits match federal: 50 mSv whole body, 150 mSv lens, 500 mSv skin/extremity, 5 mSv embryo/fetus. Fluoroscopy requires at least 0.25 mm Pb-equivalent built-in shielding; 0.35 mm aprons are required only when sterile fields or special procedures prevent barriers. Aprons are inspected annually. Engineered apron alternatives are not addressed. Operation must be by or under a licensed practitioner. The 8-hour fluoroscopically guided interventional (FGI) training rule is not enforced.

The rules, question by question

Every answer names the rule it comes from. Open "Rule text and source" to read the citation, the source page and, where it could be matched, the exact wording.

Where the rules live

Which rules apply, and whether ALARA is among them.

Is Texas an NRC Agreement State?

Yes

Yes. Texas signed its agreement on January 10, 1963, effective March 1, 1963; the NRC calls Texas the fifth Agreement State.

Rule text and source
Rule
NRC list of Agreement States, updated October 29, 2026; Agreement under Section 274 of the Atomic Energy Act
Source
nrc.gov
nrc.gov
What it says
Texas is an Agreement State.
This quote was matched word for word to the source page.

Why this matters, in ORSIF's FAQs

Which rules cover X-ray and fluoroscopy equipment in Texas?

Yes

25 TAC §289.231 (machine-only standards for protection, last amended effective Dec 29, 2011), §289.227 (healing arts X-ray and fluoroscopy, effective May 1, 2013), and §289.226 (registration). Radioactive materials fall under the separate §289.202.

Rule text and source
Rule
25 TAC §289.231(a)-(b); §289.227
Source
dshs.texas.gov
What it says
this section applies to persons who receive, possess, use, or transfer radiation machines
This quote was matched word for word to the source page.

Why this matters, in ORSIF's FAQs

Does Texas require X-ray users to keep doses as low as reasonably achievable (ALARA)?

Yes

Yes. Machine registrants must use procedures and engineering controls to keep occupational and public doses ALARA, and the operator's position during an exposure must keep the operator's dose ALARA.

Where ALARA lives in Texas: In X-ray-specific rules. ALARA appears in §289.231(l), a rule limited to radiation machines, and in §289.227(i)(10), the operator-position requirement. Radioactive materials have the separate §289.202, so an NRC-driven change there would not automatically change §289.231. §289.202 itself was not read. (25 TAC §289.231(b)(2); §289.231(l); §289.227(i)(10))

Rule text and source
Rule
25 TAC §289.231(l); §289.227(i)(10)
Source
dshs.texas.gov
What it says
The registrant shall use, to the extent practical, procedures and engineering controls based upon sound radiation protection principles
This quote was matched word for word to the source page.

Why this matters, in ORSIF's FAQs

Dose limits

The annual limits for workers, written in millisieverts (mSv). Ten mSv equal one rem.

What is the annual whole-body dose limit in Texas?

Yes

5 rem (50 mSv) TEDE per year; minors 10% of adult limits.

Rule text and source
Rule
25 TAC §289.231(m)(1)(A)-(D); §289.227(i)(3)
Source
dshs.texas.gov
What it says
An annual limit shall be the TEDE being equal to 5 rems (0.05 Sv).
This quote was matched word for word to the source page.

Why this matters, in ORSIF's FAQs

What is the annual eye-lens dose limit in Texas?

Yes

15 rem (150 mSv) lens dose equivalent (LDE) per year; Texas has not adopted the International Commission on Radiological Protection (ICRP) 20 mSv limit.

Rule text and source
Rule
25 TAC §289.231(m)(1)(A)-(D)
Source
dshs.texas.gov
What it says
an LDE of 15 rems (0.15 Sv)
This quote was matched word for word to the source page.

Why this matters, in ORSIF's FAQs

What is the annual limit for the skin, hands and feet in Texas?

Yes

50 rem (500 mSv) shallow dose equivalent (SDE) per year to the skin of the whole body or any extremity.

Rule text and source
Rule
25 TAC §289.231(m)(1)(A)-(D)
Source
dshs.texas.gov

Why this matters, in ORSIF's FAQs

What limit applies to a declared pregnant worker in Texas?

Yes

0.5 rem (5 mSv) to the embryo/fetus of a declared pregnant worker over the whole pregnancy. Large swings above a uniform monthly rate are to be avoided, citing National Council on Radiation Protection and Measurements (NCRP) Report 91's 0.5 mSv/month guidance.

Rule text and source
Rule
25 TAC §289.231(m)(1)(A)-(D)
Source
dshs.texas.gov

Why this matters, in ORSIF's FAQs

Dosimeters

When a worker must wear a dosimeter, and where.

When must a worker in Texas wear a dosimeter?

Yes

Dosimeters are required for adults likely to exceed 10% of a limit per year; minors likely to exceed 1 mSv deep dose equivalent (DDE), 1.5 mSv LDE or 5 mSv SDE; declared pregnant workers likely to exceed 1 mSv DDE over the pregnancy; and anyone entering a high or very high radiation area. Operators of only minimal-threat machines are exempt.

Rule text and source
Rule
25 TAC §289.231(n)(1)-(2)
Source
dshs.texas.gov
What it says
adults likely to receive, in 1 year from sources external to the body, a dose in excess of 10% of the limits
This quote was matched word for word to the source page.

Why this matters, in ORSIF's FAQs

Where must the dosimeter be worn in Texas when a protective apron is worn?

Yes

The badge is typically worn at the collar outside the apron and the fetal badge at the waist under the apron. Effective dose equivalent (EDE) for fluoroscopy with an apron: one collar badge counts as the reported reading, or reading × 0.3 if it exceeds 25% of the limit; with two badges, 1.5 × waist badge plus 0.04 × collar badge. The rule does not name NCRP 122 as the source.

Rule text and source
Rule
25 TAC §289.231(m)(3)-(6); §289.231(q)(1)(B)-(E)
Source
dshs.texas.gov
What it says
When a protective apron is worn, the location of the individual monitoring device is typically at the neck (collar).
This quote was matched word for word to the source page.

Why this matters, in ORSIF's FAQs

Aprons and alternatives

Protective apparel, how it is checked, and whether other systems are addressed.

Must staff in Texas fluoroscopy rooms wear protective aprons or other apparel?

Yes

The fluoroscopy system must shield the body (except head, neck and extremities) with at least 0.25 mm lead-equivalent material. Where sterile fields or special procedures prohibit normal barriers, everyone in the room except the patient must wear 0.35 mm lead-equivalent aprons. Protective devices generally must be at least 0.25 mm lead-equivalent. No thyroid shield or eyewear requirement was found.

Rule text and source
Rule
25 TAC §289.227(m)(8)(A)-(B); §289.227(i)(4)(A)
Source
dshs.texas.gov
What it says
All persons, except the patient, in the room where fluoroscopy is performed shall wear protective aprons that provide a shielding equivalent of 0.35 mm of lead.
This quote was matched word for word to the source page.

Why this matters, in ORSIF's FAQs

Does Texas require protective aprons to be inspected?

Yes

Yes. Aprons, gloves and shields must be checked annually by sight, touch or X-ray imaging. Defective items must be replaced or removed from service, and each check must be recorded.

Rule text and source
Rule
25 TAC §289.227(i)(4)(B)
Source
dshs.texas.gov
What it says
Protective devices, including aprons, gloves, and shields shall be checked annually for defects such as holes, cracks, and tears.
This quote was matched word for word to the source page.

Why this matters, in ORSIF's FAQs

Do the rules in Texas address engineered protection systems used in place of aprons?

UnclearLess certain

Not specifically addressed. The required built-in 0.25 mm shielding may be met by drapes, curtains or viewing shields, and aprons are mandatory only when sterile fields or special procedures prevent such barriers. The rules do not say whether an engineered system would suffice without aprons.

Rule text and source
Rule
25 TAC §289.227(m)(8)(A)-(B)
Source
dshs.texas.gov
What it says
The material may be, but is not limited to, drapes, self-supporting curtains, or viewing shields, in addition to any lead equivalency provided by a protective apron.
This quote was matched word for word to the source page.

About ALARA+

Who operates the equipment

Who may use fluoroscopy equipment and what training the rules ask for.

Who may operate fluoroscopy equipment in Texas?

In partLess certain

Use must be by or under supervision of a Texas-licensed physician, chiropractor or podiatrist, and operators must meet Occupations Code ch. 601 credentialing. FGI procedures are limited to radiologists, radiation oncologists, physicians and their ch. 601 delegates. Technologist, PA and nurse scope under the new TMB Chapter 186 was not verified.

Rule text and source
Rule
25 TAC §289.227(b)(1), (i)(5), (m)(9)(C)(i)(I); former 22 TAC §194.17 (repealed)
Source
dshs.texas.gov
What it says
All usage of such machines under this section shall be made by or under the supervision of a practitioner of the healing arts.
This quote was matched word for word to the source page.

Why this matters, in ORSIF's FAQs

What training or credentials does Texas require to operate fluoroscopy equipment?

In partLess certain

Non-radiologist physicians and their delegates must complete 8 hours of Category 1 continuing medical education (CME) in radiation safety plus 1 hour of hands-on training before FGI procedures, but DSHS has delayed enforcement indefinitely. Technologist certification goes through the TMB under ch. 601; details were not verified.

Rule text and source
Rule
25 TAC §289.227(m)(9)(E); DSHS §289.227 Q&A
Source
dshs.texas.gov
What it says
shall complete a minimum of 8 hours of Category 1 CMEU in radiation safety awareness training prior to performing FGI procedures
This quote was matched word for word to the source page.

Why this matters, in ORSIF's FAQs

Changes on the way, and recent ones

Rules, bills and guidance the research found from 2025 and 2026, with the ones that touch fluoroscopy teams first.

  • Adopted ruleTexas Medical Board repeal of Chapter 194 and move to Chapter 186

    Chapter 194 repealed; date not confirmed

    Technologist rules moved from 22 TAC ch. 194 to ch. 186.

    Source: law.cornell.edu

3 other items the research logged (radiation therapy, industrial machines, fees and other subjects)
  • Proposed ruleProposed amendment to §289.229 (radiation therapy machines)

    Proposed in the Texas Register (Sep 18, 2026)

    Would define "remote operation."

    Source: sos.texas.gov

  • Proposed ruleProposed amendment to §289.228 (industrial radiation machines)

    Open for comment until November 23, 2026

    The only rule listed on the DSHS proposed-rules page. No changes to §289.227 or §289.231 were found.

    Source: dshs.texas.gov

  • OtherDSHS Sunset review

    Under review

    DSHS is under Sunset review.

Also worth knowing

  • Every registrant doing FGI procedures needs a radiation protocol committee including a physician, a licensed medical physicist and the radiation safety officer (RSO), meeting at least every 14 months. It sets protocols and reference levels, and patient dose data (cumulative air kerma or DAP) must be recorded.
  • Each operator must read the written operating and safety procedures at least once a year, with documentation.
  • DSHS posts enforcement actions, including §289.227 penalties: https://www.dshs.texas.gov/sites/default/files/radiation/pdffiles/DiscEnfActions/newradiationdisciplinaryactions.pdf
  • The 8-hour FGI training rule is on the books but not enforced (delayed, per FluoroSafety, since January 13, 2015).
  • The §289.230 (mammography) revision is dated 06/06/25.
  • Texas bills on apron alternatives were not searched.

What ORSIF could not confirm

Questions the research could not settle. If you know the answer, tell us.

  • Current TMB technologist rules (22 TAC Chapter 186): who may run fluoroscopy (technologists, cardiovascular-category technologists, PAs, NPs, RNs), what supervision is needed, and any fluoroscopy, American Registry of Radiologic Technologists (ARRT) or continuing education (CE) requirements.
  • §289.202 ALARA text, and whether DSHS would carry an NRC ALARA change over to §289.231.
  • Whether the SOS official versions of §§289.227 and 289.231 match the DSHS PDFs.
  • Whether the §289.231(m)(3) badge coefficients come from NCRP Report 122.
  • Whether DSHS would accept an engineered shielding system under §289.227(m)(8)(A), or would require a variance or exemption.
  • 2025-2026 Texas bills on apron alternatives or fluoroscopy operators (check https://capitol.texas.gov).
  • Current enforcement status of the FGI training rule; the DSHS delay letter was unreadable.

Who regulates, and where to read the rules

Documents ORSIF has on file

Copies saved earlier. Rules change, so check the current text at the source.

  • 25 TAC §289.227 - X-rays in Healing Arts Download
  • 25 TAC §289.231 - General Radiation Control Download
  • RC203-1 - Notices to Workers Download

Double-check before you rely on this page

ORSIF compiled this page from the sources linked on it and did its best. It is not comprehensive and it is not always accurate. Rules change, a rule may have been missed or misread, and the sources themselves can be out of date. Treat this page as a starting point.

Where to check

What may be missing here

  • 1 of 14 answers on this page are "Not found" or "Unclear". That means ORSIF did not find the answer in the rules it reviewed, not that no rule exists.
  • 3 answers carry a flag or a lower confidence ("Check the source" or "Less certain").
  • Researched November 3, 2026. Rules may have changed since.
  • Only the state's rules for X-ray and fluoroscopy equipment were reviewed.

Found something wrong or missing? Tell ORSIF.

Sources and terms

Terms used on this page
ALARA
As low as reasonably achievable: keep radiation doses as low as is reasonable, below the legal limits.
Agreement State
A state that has signed an agreement with the U.S. Nuclear Regulatory Commission to regulate certain radioactive materials itself.
mSv
Millisievert, a unit of radiation dose. Ten mSv equal one rem, the older unit some rules still use.
Dosimeter
A badge or device worn to measure a worker's radiation dose.
Lead equivalent
The thickness of lead (mm Pb) that would give the same protection as the material described.
NRC
U.S. Nuclear Regulatory Commission
ICRP
International Commission on Radiological Protection
NCRP
National Council on Radiation Protection and Measurements
TEDE
Total effective dose equivalent
DDE
Deep dose equivalent
LDE
Lens dose equivalent
SDE
Shallow dose equivalent
EDE
Effective dose equivalent
ARRT
American Registry of Radiologic Technologists
FGI
Fluoroscopically guided interventional
RSO
Radiation safety officer
CME
Continuing medical education
CE
Continuing education
All 16 sources the research read