Please double-check this information. This is an ORSIF research summary, researched November 3, 2026. It is not comprehensive and not always accurate: rules change, and answers may be incomplete, out of date or wrong. It is education, not legal advice. Not yet reviewed by a regulatory expert. Confirm each requirement at the source linked beside the answer, with the state program and with your facility's radiation safety officer. How this was made

In brief

ORSIF's research identified the Rhode Island Department of Health (RIDOH), Center for Health Facilities Regulation, Radiation Control Program ("Rhode Island Radiation Control Agency (RCA)") as the agency that regulates X-ray and fluoroscopy equipment in Rhode Island. Rhode Island is a U.S. Nuclear Regulatory Commission (NRC) Agreement State: it has signed an agreement with the NRC to regulate certain radioactive materials itself.

Rhode Island (an Agreement State) has no ALARA wording of its own for X-ray users. ALARA applies through 216-RICR-40-20-1, which adopts the fixed 2021 edition of 10 CFR Part 20 and deems "licensee" to include "registrant," so the requirement sits in a rule shared with radioactive-materials licensees. Dose limits are the federal ones: 50 mSv whole body, 150 mSv lens of the eye, 500 mSv skin and extremities, and 5 mSv embryo/fetus. Everyone in the fluoroscopy room except the patient must wear 0.25 mm lead-equivalent garments where sterile fields prevent the use of drapes or barriers. Aprons must be checked for integrity annually. The general X-ray rule accepts "whole-body protective barriers," but engineered apron replacements are never mentioned. Fluoroscopy is operated by licensed radiographers or by exempt practitioners. All fluoroscopy operators and supervisors need 4 hours of training (8 for interventional work) plus refreshers every 2 years.

The rules, question by question

Every answer names the rule it comes from. Open "Rule text and source" to read the citation, the source page and, where it could be matched, the exact wording.

Where the rules live

Which rules apply, and whether ALARA is among them.

Is Rhode Island an NRC Agreement State?

Yes

Yes. Rhode Island is not on the NRC list of non-Agreement States (updated 2026-10-29), so it is one of the 41 Agreement States.

Rule text and source
Rule
NRC list of Agreement States, updated October 29, 2026
Source
nrc.gov

Why this matters, in ORSIF's FAQs

Which rules cover X-ray and fluoroscopy equipment in Rhode Island?

YesCheck the source

Yes. The X-ray/fluoroscopy rule is 216-RICR-40-20-4, Medical Diagnostic and Interventional X-Ray and Imaging Systems. Fluoroscopy is in §4.5 and training in §4.3.3. Per Cornell LII, §4.5 was amended effective 11/28/2023. Part 4 adopts the ACR-AAPM fluoroscopy standard (2021) and National Council on Radiation Protection and Measurements (NCRP) 168 by reference. The general standards rule is 216-RICR-40-20-1, which applies to licensees and registrants and adopts 10 CFR Part 20 (2021) by reference; its effective date is unconfirmed, possibly 5/25/2022. Facility registration is in 216-RICR-40-20-3.

Rule text and source
Rule
216-RICR-40-20-4 §4.1(B); 216-RICR-40-20-1 §1.1(B), §1.2(A)
Source
rules.sos.ri.gov

Why this matters, in ORSIF's FAQs

Does Rhode Island require X-ray users to keep doses as low as reasonably achievable (ALARA)?

Yes

Yes, but only indirectly. Part 1 adopts 10 CFR Part 20 (2021), including 20.1101(b) (ALARA), and deems "licensee" to include "registrant."

Where ALARA lives in Rhode Island: Only in a rule shared with radioactive materials. ALARA reaches X-ray registrants only through Part 1's adoption of the fixed 2021 edition of 10 CFR Part 20, which is shared with radioactive-materials licensees. NRC changes do not flow in automatically; RIDOH would have to amend Part 1. Part 4 has no ALARA wording; §4.3.10(B), which requires minimizing exposure, is the closest language, and the research does not treat it as an ALARA requirement. (216-RICR-40-20-1 §1.2; 216-RICR-40-20-4 §4.3.10(B))

Rule text and source
Rule
216-RICR-40-20-1 §1.2(A)-(C)(5), §1.6; 10 CFR 20.1101(b) (2021 edition adopted)
Source
rules.sos.ri.gov
What it says
Any reference to licensee shall be deemed to include registrant.
This quote was matched word for word to the source page.

Why this matters, in ORSIF's FAQs

Dose limits

The annual limits for workers, written in millisieverts (mSv). Ten mSv equal one rem.

What is the annual whole-body dose limit in Rhode Island?

Yes

5 rem (50 mSv) total effective dose equivalent per year, via 10 CFR 20.1201 adopted by reference.

Rule text and source
Rule
216-RICR-40-20-1 §1.7.1(A); 10 CFR 20.1201(a)
Source
rules.sos.ri.gov
What it says
For the purpose of this Part, the occupational dose limits for adults are defined by 10 C.F.R. § 20.1201.
This quote was matched word for word to the source page.

Why this matters, in ORSIF's FAQs

What is the annual eye-lens dose limit in Rhode Island?

Yes

15 rem (150 mSv) per year. No lower lens limit has been adopted.

Rule text and source
Rule
216-RICR-40-20-1 §1.7.1(A); 10 CFR 20.1201(a)(2)(i)
Source
govinfo.gov
What it says
A lens dose equivalent of 15 rems (0.15 Sv), and
This quote was matched word for word to the source page.

Why this matters, in ORSIF's FAQs

What is the annual limit for the skin, hands and feet in Rhode Island?

Yes

50 rem (500 mSv) shallow-dose equivalent to the skin and extremities per year.

Rule text and source
Rule
216-RICR-40-20-1 §1.7.1(A); 10 CFR 20.1201(a)
Source
ecfr.gov

Why this matters, in ORSIF's FAQs

What limit applies to a declared pregnant worker in Rhode Island?

Yes

0.5 rem (5 mSv) to the embryo/fetus of a declared pregnant worker for the whole pregnancy.

Rule text and source
Rule
216-RICR-40-20-1 §1.7.8; 10 CFR 20.1208(a)
Source
ecfr.gov

Why this matters, in ORSIF's FAQs

Dosimeters

When a worker must wear a dosimeter, and where.

When must a worker in Rhode Island wear a dosimeter?

Yes

Monitoring is required for adults likely to exceed 10% of the adult limits in a year, declared pregnant women likely to exceed 1 mSv (0.1 rem) deep dose during the pregnancy, minors above lower thresholds, and anyone entering a high or very high radiation area.

Rule text and source
Rule
216-RICR-40-20-1 §1.10.3(A); 10 CFR 20.1502(a)
Source
ecfr.gov
What it says
Adults likely to receive, in 1 year from sources external to the body, a dose in excess of 10 percent of the limits in § 20.1201(a),
This quote was matched word for word to the source page.

Why this matters, in ORSIF's FAQs

Where must the dosimeter be worn in Rhode Island when a protective apron is worn?

Yes

Yes, state-specific rules apply. A single badge is worn at the neck outside the apron; if the reported dose is above 25% of the limit, effective dose equivalent = deep dose × 0.3. With two badges (waist under the apron, neck outside it), effective dose equivalent = 1.5 × waist + 0.04 × neck. A declared pregnant woman needs a waist badge under the apron; if it reads more than 0.5 mSv in a month, a medical physicist should be consulted. The eye-dose badge goes at the neck or an unshielded spot closer to the eye. The extremity badge goes on the extremity likely to receive the highest dose.

Rule text and source
Rule
216-RICR-40-20-1 §1.7.1(B)(1)-(2); §1.10.3(B)-(C)
Source
rules.sos.ri.gov
What it says
When only one individual monitoring device is used to determine the effective dose equivalent for external radiation pursuant to § 1.7.1(B) of this Part, it shall be located at the neck outside the protective apron.
This quote was matched word for word to the source page.

Why this matters, in ORSIF's FAQs

Aprons and alternatives

Protective apparel, how it is checked, and whether other systems are addressed.

Must staff in Rhode Island fluoroscopy rooms wear protective aprons or other apparel?

Yes

Yes. Stationary fluoroscopy systems need drapes, curtains or shields, used for everyone except the patient. Where sterile fields prevent their use, everyone except the patient must wear garments of at least 0.25 mm lead equivalent. Anyone in the direct beam needs 0.5 mm. Scatter protection requires 0.25 mm garments or whole-body barriers. Facilities must keep enough aprons, gloves and collars available. No thyroid shield or eyewear requirement was found.

Rule text and source
Rule
216-RICR-40-20-4 §4.5.12(A)-(B); §4.3.6(A)(1)-(2); §4.3.9(E)
Source
rules.sos.ri.gov
What it says
All persons, except the patient, in the room where fluoroscopy is performed shall wear protective garments that provide a lead equivalent shielding of at least twenty-five one hundredths of one millimeter (0.25 mm);
This quote was matched word for word to the source page.

Why this matters, in ORSIF's FAQs

Does Rhode Island require protective aprons to be inspected?

Yes

Yes. All protective apparel and auxiliary shields must be evaluated annually for integrity and labeled with their lead equivalence. The method is not specified.

Rule text and source
Rule
216-RICR-40-20-4 §4.3.9(E)(1)
Source
rules.sos.ri.gov
What it says
All protective apparel and auxiliary shields shall be evaluated annually for integrity and clearly labeled with their lead equivalence.
This quote was matched word for word to the source page.

Why this matters, in ORSIF's FAQs

Do the rules in Rhode Island address engineered protection systems used in place of aprons?

UnclearLess certain

Not addressed by name. Engineered systems that replace aprons are never mentioned. §4.3.6(A)(2) accepts "whole-body protective barriers" as an alternative to garments for scatter protection, and §1.5.2(A) allows Agency exemptions. The incorporated ACR-AAPM and NCRP documents were not read.

Rule text and source
Rule
216-RICR-40-20-4 §4.3.6(A)(2), §4.5.12(A); 216-RICR-40-20-1 §1.5.2(A)
Source
rules.sos.ri.gov
What it says
All individuals shall be protected from the secondary radiation by protective garments or whole-body protective barriers of not less than twenty-five one hundredths of one quarter millimeter (0.25 mm) lead equivalent material.
This quote was matched word for word to the source page.

About ALARA+

Who operates the equipment

Who may use fluoroscopy equipment and what training the rules ask for.

Who may operate fluoroscopy equipment in Rhode Island?

Yes

Licensed (ARRT-certified) radiographers under the general supervision of a licensed practitioner. Exempt from the technologist license: licensed practitioners within their field (physicians, chiropractors, podiatrists, nurse practitioners, physician assistants), American Registry of Radiologic Technologists (ARRT) radiologist assistants, supervised medical students, medical physicists and supervised radiologic technology students. For interventional (FGI) procedures, the Radiation Protocol Committee identifies authorized users. No fluoroscopy-specific supervision rule was found.

Rule text and source
Rule
216-RICR-40-20-4 §4.1(B), §4.3.3(A), §4.5.13(F)(1); R.I. Gen. Laws §§ 5-68.1-2(7), 5-68.1-4(a); 216-RICR-40-05-34 §34.2(A)(14), (20)
Source
webserver.rilegislature.gov
What it says
A licensed practitioner when practicing within his or her field of expertise.
This quote was matched word for word to the source page.

Why this matters, in ORSIF's FAQs

What training or credentials does Rhode Island require to operate fluoroscopy equipment?

Yes

Everyone who operates or supervises fluoroscopy, physicians included, needs at least 4 hours of training before performing fluoroscopy. Interventional (FGI) work needs 8 hours of Agency-approved training, including 1 hour hands-on. Training is given by a Qualified Medical Physicist or an Agency-approved individual. Refreshers are required every 24 months (2 hours in-service or continuing medical education (CME)), and records are kept for 3 years. Radiographers need ARRT certification. There is no separate fluoroscopy permit, and no continuing-education requirement for license renewal was found.

Rule text and source
Rule
216-RICR-40-20-4 §4.3.3(C)-(G); 216-RICR-40-05-34 §34.4(A)
Source
rules.sos.ri.gov
What it says
All individuals operating, or supervising the operation of, fluoroscopic X-ray systems shall have completed a minimum of four (4) hours training, prior to performing fluoroscopy procedures
This quote was matched word for word to the source page.

Why this matters, in ORSIF's FAQs

Also worth knowing

  • Facilities doing interventional (FGI) procedures must have a Radiation Protocol Committee (physician, Qualified Medical Physicist, lead technologist) that meets at least every 12 months and sets protocols, dose notification levels and substantial radiation dose levels (§4.5.13).
  • Part 4 adopts the ACR-AAPM fluoroscopy standard (2021) and NCRP Report 168 (2010) by reference (§4.1.1).
  • No individual may be used routinely to hold patients or image receptors; holders must be instructed and protected (§4.3.9).
  • Only people required for the procedure (plus trainees, or a parent or guardian) may be near the patient while the tube is energized (§4.3.6(A)).
  • Prior occupational dose must be recorded on Agency Form RCA-2 or equivalent (§1.7.5).
  • Deliberately exposing a badge to give a false reading is prohibited (§1.10.2(B)).
  • Pending changes for 2025-2026 were not found. Part 4 was last amended effective 11/28/2023, Part 34 was refiled effective 01/04/2022, and Part 1 has a 2022 version. RIDOH Regulatory News (about March-May 2026) listed no radiation-rule items, and one search found no bill on apron alternatives. The search was not exhaustive.

What ORSIF could not confirm

Questions the research could not settle. If you know the answer, tell us.

  • Confirm that the 2021 edition of 10 CFR Part 20 matches the current eCFR and 2025 GovInfo text that was read.
  • Confirm Part 1's effective date (probably May 2022).
  • Search fully for 2025-2026 rulemaking and bills, and ask whether RIDOH plans to follow the NRC's "graded approach" change.
  • Check whether the incorporated ACR-AAPM (2021) standard and NCRP Report 168 contain binding staff-protection, eyewear or shielding provisions.
  • Clarify how §4.3.6 applies to fluoroscopy compared with §4.5.12, and whether "whole-body protective barriers" or a §1.5.2 exemption would cover an engineered apron-replacement system.
  • Confirm whether any continuing-education requirement applies to radiographer license renewal.

Double-check before you rely on this page

ORSIF compiled this page from the sources linked on it and did its best. It is not comprehensive and it is not always accurate. Rules change, a rule may have been missed or misread, and the sources themselves can be out of date. Treat this page as a starting point.

Where to check

What may be missing here

  • 1 of 14 answers on this page are "Not found" or "Unclear". That means ORSIF did not find the answer in the rules it reviewed, not that no rule exists.
  • 2 answers carry a flag or a lower confidence ("Check the source" or "Less certain").
  • Researched November 3, 2026. Rules may have changed since.
  • Only the state's rules for X-ray and fluoroscopy equipment were reviewed.

Found something wrong or missing? Tell ORSIF.

Sources and terms

Terms used on this page
ALARA
As low as reasonably achievable: keep radiation doses as low as is reasonable, below the legal limits.
Agreement State
A state that has signed an agreement with the U.S. Nuclear Regulatory Commission to regulate certain radioactive materials itself.
mSv
Millisievert, a unit of radiation dose. Ten mSv equal one rem, the older unit some rules still use.
Dosimeter
A badge or device worn to measure a worker's radiation dose.
Lead equivalent
The thickness of lead (mm Pb) that would give the same protection as the material described.
NRC
U.S. Nuclear Regulatory Commission
NCRP
National Council on Radiation Protection and Measurements
ARRT
American Registry of Radiologic Technologists
CME
Continuing medical education
All 16 sources the research read