Is Oregon an NRC Agreement State?
Yes. This is per the NRC list, and Oregon's rule text confirms it.
Rule text and source
- Rule
- NRC list of Agreement States, updated October 29, 2026; OAR 333-120-0000(1)
- Source
- nrc.gov
secure.sos.state.or.us
What Oregon's rules say about ALARA (as low as reasonably achievable), dose limits in millisieverts (mSv), dosimeters, aprons and training for the people who work in fluoroscopy labs.
Please double-check this information. This is an ORSIF research summary, researched November 3, 2026, with its weakest answers rechecked November 5, 2026. It is not comprehensive and not always accurate: rules change, and answers may be incomplete, out of date or wrong. It is education, not legal advice. Not yet reviewed by a regulatory expert. Confirm each requirement at the source linked beside the answer, with the state program and with your facility's radiation safety officer. How this was made
ORSIF's research identified the Oregon Health Authority (OHA), Public Health Division, Center for Health Protection, Radiation Protection Services (RPS), X-ray program as the agency that regulates X-ray and fluoroscopy equipment in Oregon. Oregon is a U.S. Nuclear Regulatory Commission (NRC) Agreement State: it has signed an agreement with the NRC to regulate certain radioactive materials itself.
Oregon's X-ray and radioactive-materials rules share division OAR 333-120, where the ALARA requirement applies to each licensee or registrant, and the X-ray rules (OAR 333-106) add a separate "minimize" duty. Dose limits include 50 mSv per year for the whole body and 150 mSv per year for the lens of the eye. Staff need aprons or whole-body protective barriers of at least 0.25 mm lead equivalent, and aprons must be inspected annually; engineered radiation protection systems are not mentioned by name, but in fluoroscopy, scatter must be blocked by an additional 0.25 mm lead equivalent, such as drapes, panels or curtains, beyond the apron unless enough distance is kept. Fluoroscopy may be operated by radiologists, trained non-radiologist physicians, OBMI-licensed technologists, permitted physician associates and radiology assistants, and PH 28-2026 tightened training for non-radiologist physicians.
Every answer names the rule it comes from. Open "Rule text and source" to read the citation, the source page and, where it could be matched, the exact wording.
Which rules apply, and whether ALARA is among them.
Yes. This is per the NRC list, and Oregon's rule text confirms it.
X-ray and fluoroscopy rules are in OAR chapter 333, division 106, "X-Rays in the Healing Arts" (fluoroscopy at 333-106-0201 to -0245). General standards for both licensees and registrants are in OAR 333-120, "Standards for Protection Against Radiation." Division 106 applies in addition to other divisions. Registration rules are in divisions 100 and 101, which were not reviewed. The fetched division pages were cut off: division 106 at about 0360 and division 120 at about 0630.
This division establishes standards for protection against ionizing radiation resulting from activities conducted under licenses and registrations issued by the Authority.This quote was matched word for word to a page the research read.
Yes. Each licensee or registrant must use procedures and engineering controls to keep occupational doses ALARA, and must review the radiation protection program at least once a year. The X-ray rules add a separate duty to minimize patient and personnel exposure (333-106-0045(1)). ALARA is also a required topic in operator safety training.
Where ALARA lives in Oregon: Only in a rule shared with radioactive materials. The ALARA requirement sits in shared division 120 (333-120-0020), which applies to licensees and registrants alike. Division 106 has no ALARA requirement of its own, only a separate "minimize" duty and ALARA as a training topic. A rewrite of 333-120-0020 would likely change the explicit ALARA duty for X-ray registrants; the research calls this an observation, not a legal conclusion. (OAR 333-120-0020(2); 333-120-0015(8); 333-106-0045(1); 333-106-0050)
Each licensee or registrant must use, to the extent practicable, procedures and engineering controls based upon sound radiation protection principles to achieve occupational doses and doses to members of the public that are as low as is reasonably achievable (ALARA).This quote was matched word for word to a page the research read.
The annual limits for workers, written in millisieverts (mSv). Ten mSv equal one rem.
50 mSv (5 rem) total effective dose equivalent per year. Any organ or tissue except the lens: 500 mSv. Minors: 10% of the adult limits.
The sum of the deep-dose equivalent and the committed dose equivalent to any individual organ or tissue other than the lens of the eye being equal to 0.5 Sv (50 rem).This quote was matched word for word to a page the research read.
150 mSv (15 rem) per year. Oregon has not adopted a lower limit such as the International Commission on Radiological Protection (ICRP)'s 20 mSv.
(A) A lens dose equivalent of 0.15 Sv (15 rem); and (B) A shallow-dose equivalent of 0.50 Sv (50 rem) to the skin of the whole body or to the skin of any extremity.This quote was matched word for word to a page the research read.
Shallow dose of 500 mSv (50 rem) per year to the skin or any extremity.
(A) A lens dose equivalent of 0.15 Sv (15 rem); and (B) A shallow-dose equivalent of 0.50 Sv (50 rem) to the skin of the whole body or to the skin of any extremity.This quote was matched word for word to a page the research read.
5 mSv (0.5 rem) to the embryo/fetus of a declared pregnant woman over the whole pregnancy.
The licensee or registrant must ensure that the dose equivalent to an embryo/fetus during the entire pregnancy, due to occupational exposure of a declared pregnant woman does not exceed five mSv (0.5 rem).This quote was matched word for word to a page the research read.
When a worker must wear a dosimeter, and where.
A dosimeter must be worn by: adults likely to exceed 10% of an adult limit in a year; minors likely to exceed 10% of a minor's limit; declared pregnant women likely to exceed 1 mSv (0.1 rem) over the pregnancy; and anyone entering a high or very high radiation area. This applies to X-ray users through 333-106-0050.
(a) Adults likely to receive, in one year from sources external to the body, a dose in excess of 10 percent of the limits in OAR 333-120-0100(1);This quote was matched word for word to a page the research read.
When an apron is worn, at least one badge goes at the collar outside the apron. A lens badge goes at the neck or closer to the eye, outside the apron. A second badge goes under the apron at the waist, and is required for a declared pregnant woman. Effective dose for fluoroscopy: (A) one neck badge, the deep dose counts as the effective dose; (B) one neck badge above 25% of the limit, deep dose × 0.3; (C) two badges, 1.5 × waist badge + 0.04 × neck badge. Other methods may be used if approved by the Authority. National Council on Radiation Protection and Measurements (NCRP) 122 and Webster are not named. The older 333-106-0050(1)(b) "most critical organ" language may be inconsistent with this.
the sum of the deep dose equivalent reported for the individual monitoring device located at the waist under the protective apron multiplied by 1.5 and the deep dose equivalent reported for the individual monitoring device located at the neck outside the protective apron multiplied by 0.04.This quote was matched word for word to a page the research read.
Protective apparel, how it is checked, and whether other systems are addressed.
Staff must be protected from scatter by aprons or whole-body protective barriers of at least 0.25 mm lead equivalent. Any body part in the useful beam needs 0.5 mm lead equivalent. In fluoroscopy rooms, under-table scatter must be blocked by at least 0.25 mm lead equivalent. For above-table scatter, staff must be at least 120 cm from the beam center, or the scatter must pass through an additional 0.25 mm lead equivalent beyond the apron; exceptions may be granted on application. Thyroid shields and eyewear are not required for staff and appear only as training topics.
Staff and ancillary personnel shall be protected from the direct scatter radiation by protective aprons or whole body protective barriers of not less than 0.25 mm lead equivalent.This quote was matched word for word to a page the research read.
Yes. Aprons, gloves and shields must be checked annually for defects, with records kept. Defective items must be replaced or removed from service. Fluoroscopy may be used to check an item only if a visual and manual check suggests a problem.
Protective equipment including aprons, gloves and shields shall be checked annually for defects, such as holes, cracks and tears to assure reliability and integrity.This quote was matched word for word to a page the research read.
The rules do not name engineered shielding systems. However, staff may be protected from scatter by either protective aprons or whole-body protective barriers of at least 0.25 mm lead equivalent. In fluoroscopy, staff must either stand at least 120 cm from the center of the beam or be shielded by at least 0.25 mm lead equivalent material such as drapes, panels or curtains in addition to the apron, and the state may grant exceptions on request.
Staff and ancillary personnel shall be protected from the direct scatter radiation by protective aprons or whole body protective barriers of not less than 0.25 mm lead equivalent.This quote was matched word for word to a page the research read.
Who may use fluoroscopy equipment and what training the rules ask for.
Radiologists, with no supervision. Trained non-radiologist practitioners (MD, DO, DC, DPM, DVM), without supervision if the registrant arranges radiologist or physicist involvement. OBMI-licensed radiologic technologists, under radiologist personal or direct supervision, trained non-radiologist personal supervision, or personal supervision by an APRN with an OBMI Limited Fluoroscopy Supervision Permit. Physician associates with an OBMI permit and a collaborative agreement, with no supervision. Radiology assistants, under direct radiologist supervision. Students, under personal supervision. Trained veterinary operators. APRNs and nurses are not listed as operators.
An advanced practice registered nurse (APRN) who has obtained an OBMI Limited Fluoroscopy Supervision Permit may provide personal supervision to radiologic technologists who have a permanent or temporary license issued from the OBMI to operate fluoroscopic equipment.This quote was matched word for word to a page the research read.
All operators need training on a set curriculum, including dose reduction, protective devices, dosimeter placement and dose limits. OBMI-licensed radiographers, permitted physician associates, radiology assistants and radiologists are deemed to meet it. Non-radiologist practitioners need OHA-approved training unless they were operating before April 11, 2005; since PH 28-2026 this also applies when they supervise an operator in the room. No fluoroscopy training hours or continuing education were found in the RPS rules. OBMI rules were not reviewed.
(b) Non-radiologist practitioners who have successfully completed a training program from an Oregon Health Authority (Authority) approved resource that meets section (2) of this rule, or have been operating fluoroscopic equipment prior to April 11, 2005;This quote was matched word for word to a page the research read.
Rules, bills and guidance the research found from 2025 and 2026, with the ones that touch fluoroscopy teams first.
Filed and effective 07/02/2026 (notice filed 04/29/2026; comment closed May 21, 2026) (Jul 2, 2026)
Amended 333-106-0005 so supervision includes the duty to direct the work. Amended 333-106-0205 to require OHA-approved training for non-radiologist practitioners who operate fluoroscopy or supervise an operator in the room. Added rules for veterinary operators.
Lead only; seen as a search excerpt and not read (Feb 11, 2026)
The slides mention a preliminary legislative concept on 2025-2031 funding.
Questions the research could not settle. If you know the answer, tell us.
ORSIF compiled this page from the sources linked on it and did its best. It is not comprehensive and it is not always accurate. Rules change, a rule may have been missed or misread, and the sources themselves can be out of date. Treat this page as a starting point.
Found something wrong or missing? Tell ORSIF.
Found something out of date or wrong? Tell ORSIF. Corrections are checked against the rule text and the page is updated.