Please double-check this information. This is an ORSIF research summary, researched November 3, 2026. It is not comprehensive and not always accurate: rules change, and answers may be incomplete, out of date or wrong. It is education, not legal advice. Not yet reviewed by a regulatory expert. Confirm each requirement at the source linked beside the answer, with the state program and with your facility's radiation safety officer. How this was made

In brief

ORSIF's research identified the Ohio Department of Health, Bureau of Environmental Health and Radiation Protection (BEHRP), X-ray Program as the agency that regulates X-ray and fluoroscopy equipment in Ohio. Ohio is a U.S. Nuclear Regulatory Commission (NRC) Agreement State: it has signed an agreement with the NRC to regulate certain radioactive materials itself.

Ohio requires X-ray and fluoroscopy registrants to keep doses ALARA under its general radiation protection chapter (OAC 3701:1-38), which is shared with radioactive-materials licensees. The fluoroscopy rules only mention ALARA reviews by hospital quality assurance (QA) committees. Dose limits match NRC levels: 50 mSv whole body, 150 mSv lens, 500 mSv skin and extremities, and 5 mSv over a declared pregnancy. Everyone in the fluoroscopy room must use a whole-body barrier or an apron of at least 0.25 mm lead equivalent, and aprons must be checked annually. Engineered alternatives to aprons are not addressed. Physicians are exempt from radiologic licensing, and licensed radiographers' scope includes fluoroscopy. All fluoroscopy operators need 2 hours of fluoroscopy-specific training. ODH posted draft amendments on 2026-09-08, with comments due 2026-10-08; their contents were not reviewed.

The rules, question by question

Every answer names the rule it comes from. Open "Rule text and source" to read the citation, the source page and, where it could be matched, the exact wording.

Where the rules live

Which rules apply, and whether ALARA is among them.

Is Ohio an NRC Agreement State?

Yes

Yes. Ohio is an NRC Agreement State, based on the NRC list (updated 2026-10-29), which does not name Ohio among non-Agreement States. OAC 3701:1-38-07(A) refers to equivalent NRC or agreement state regulations.

Rule text and source
Rule
NRC list of Agreement States, updated October 29, 2026; OAC 3701:1-38-07(A).
Source
nrc.gov

Why this matters, in ORSIF's FAQs

Which rules cover X-ray and fluoroscopy equipment in Ohio?

Yes

General standards are in OAC Chapter 3701:1-38, which covers both radioactive-materials licensees and X-ray registrants (key rules -07, -11, -12, -14). X-ray and fluoroscopy rules are in OAC Chapter 3701:1-66 (-01, -02, -04, -07). Operator licensing is in OAC Chapter 3701-72 and ORC Chapter 4773.

Rule text and source
Rule
OAC 3701:1-38-07(A); OAC 3701:1-66-02 (introductory paragraph).
Source
codes.ohio.gov
What it says
The rules in Chapter 3701:1-38 of the Administrative Code apply to facilities that receive, possess, use, transfer, or dispose of radioactive material, and to handlers of radiation-generating equipment.
This quote was matched word for word to the source page.

Why this matters, in ORSIF's FAQs

Does Ohio require X-ray users to keep doses as low as reasonably achievable (ALARA)?

Yes

Yes. Licensees and registrants must use procedures and engineering controls to keep occupational and public doses ALARA. They must also have a documented radiation protection program and review it at least every 12 months. The air-emissions dose constraint applies only to licensees.

Where ALARA lives in Ohio: Only in a rule shared with radioactive materials. The binding ALARA duty is in 3701:1-38-11(D)(2), in a chapter that applies to both radioactive-materials licensees and radiation-generating equipment handlers. Chapter 3701:1-66 only mentions ALARA reviews in hospital QA committee minutes (3701:1-66-04(C)(4)(c)) and does not restate the standard. (OAC 3701:1-38-07(A), 3701:1-38-11(D)(2), 3701:1-66-04(C)(4)(c).)

Rule text and source
Rule
OAC 3701:1-38-11(D)(2) (effective 2011-09-01).
Source
codes.ohio.gov
What it says
The licensee or registrant shall use, to the extent practicable, procedures and engineering controls based upon sound radiation protection principles to achieve occupational doses and public doses that are as low as is reasonably achievable.
This quote was matched word for word to the source page.

Why this matters, in ORSIF's FAQs

Dose limits

The annual limits for workers, written in millisieverts (mSv). Ten mSv equal one rem.

What is the annual whole-body dose limit in Ohio?

Yes

0.05 Sv (5 rem) = 50 mSv per year TEDE.

Rule text and source
Rule
OAC 3701:1-38-12(A)(1)-(2), (G), (H) (effective 2010-10-04; five-year review due 2027-01-20).
Source
codes.ohio.gov

Why this matters, in ORSIF's FAQs

What is the annual eye-lens dose limit in Ohio?

Yes

0.15 Sv (15 rem) = 150 mSv per year. Ohio has not adopted a lower limit such as 20 mSv.

Rule text and source
Rule
OAC 3701:1-38-12(A)(1)-(2), (G), (H) (effective 2010-10-04; five-year review due 2027-01-20).
Source
codes.ohio.gov
What it says
(a) A lens dose equivalent of 0.15 sievert (fifteen rem), and (b) A shallow-dose equivalent of 0.5 sievert (fifty rem) to the skin of the whole body or to the skin of any extremity.
This quote was matched word for word to the source page.

Why this matters, in ORSIF's FAQs

What is the annual limit for the skin, hands and feet in Ohio?

Yes

0.5 Sv (50 rem) = 500 mSv per year to the skin of the whole body or any extremity.

Rule text and source
Rule
OAC 3701:1-38-12(A)(1)-(2), (G), (H) (effective 2010-10-04; five-year review due 2027-01-20).
Source
codes.ohio.gov
What it says
(a) A lens dose equivalent of 0.15 sievert (fifteen rem), and (b) A shallow-dose equivalent of 0.5 sievert (fifty rem) to the skin of the whole body or to the skin of any extremity.
This quote was matched word for word to the source page.

Why this matters, in ORSIF's FAQs

What limit applies to a declared pregnant worker in Ohio?

Yes

5 mSv (0.5 rem) to the embryo/fetus of a declared pregnant woman over the whole pregnancy. If an apron is worn, the fetal dose is taken from the mother's lower torso region (3701:1-38-12(H)(4)).

Rule text and source
Rule
OAC 3701:1-38-12(A)(1)-(2), (G), (H) (effective 2010-10-04; five-year review due 2027-01-20).
Source
codes.ohio.gov

Why this matters, in ORSIF's FAQs

Dosimeters

When a worker must wear a dosimeter, and where.

When must a worker in Ohio wear a dosimeter?

Yes

A badge must be supplied and worn by: adults likely to receive more than 10% of an annual limit (5 mSv TEDE, 15 mSv lens, 50 mSv skin or extremity); minors above set levels; declared pregnant women likely to receive more than 1 mSv over the pregnancy; and anyone entering high or very high radiation areas. Anyone taking part in fluoroscopic procedures must wear a monitoring device unless the IRRP or a radiation expert determines those thresholds are unlikely to be exceeded.

Rule text and source
Rule
OAC 3701:1-38-14(B)(1); OAC 3701:1-66-02(G)(5).
Source
codes.ohio.gov
What it says
Any radiation worker participating in fluoroscopic, veterinary, or mobile or portable x-ray procedures is obligated to wear an individual monitoring device unless the IRRP or radiation expert determines it is unlikely the radiation worker will receive in excess of the doses specified...
This quote was matched word for word to the source page.

Why this matters, in ORSIF's FAQs

Where must the dosimeter be worn in Ohio when a protective apron is worn?

Yes

With an apron, the whole-body badge is typically worn at the neck. The lens badge goes at the neck outside the apron or near the eye. A fetal badge goes at the waist under the apron, and a second badge is required for a declared pregnant woman. Effective dose is calculated as follows: one neck badge outside the apron × 0.3; with two badges, (waist badge under the apron × 1.5) + (neck badge outside the apron × 0.04); or ANSI/National Council on Radiation Protection and Measurements (NCRP)/director-approved methods. Badges must be processed by a National Voluntary Laboratory Accreditation Program (NVLAP)-accredited processor.

Rule text and source
Rule
OAC 3701:1-38-12(A)(4)(b); OAC 3701:1-38-14(C)(1)-(5).
Source
codes.ohio.gov
What it says
When only one individual monitoring device is used and it is located at the neck outside the protective apron, the reported deep dose equivalent value multiplied by 0.3 shall be the effective dose equivalent for external radiation
This quote was matched word for word to the source page.

Why this matters, in ORSIF's FAQs

Aprons and alternatives

Protective apparel, how it is checked, and whether other systems are addressed.

Must staff in Ohio fluoroscopy rooms wear protective aprons or other apparel?

Yes

Anyone in the room during fluoroscopy must stand behind a whole-body protective barrier or wear an apron of at least 0.25 mm lead equivalent. This may be skipped at a distance and for a time a radiation expert has documented in the QA program as unlikely to exceed 2 mrem in any hour or 100 mrem a year. Lead gloves are required for hands in the useful beam, and 0.5 mm lead equivalent for any body part in the useful beam. Enough aprons and gloves must be available. No thyroid shield or eyewear requirement was found. A 2025 PMC review lists Ohio as requiring 0.50 mm, which conflicts with the primary text.

Rule text and source
Rule
OAC 3701:1-66-07(B)(1)-(2) (effective 2024-12-13); OAC 3701:1-66-02(G)(2), (G)(4).
Source
codes.ohio.gov
What it says
Any individual who is in the room during the fluoroscopic procedure will be adequately protected by standing behind a whole body protective barrier or will wear a protective lead apron of not less than 0.25 millimeter lead equivalent.
This quote was matched word for word to the source page.

Why this matters, in ORSIF's FAQs

Does Ohio require protective aprons to be inspected?

Yes

Yes. An annual evaluation of the integrity of all necessary protective apparel is required. The method is not specified.

Rule text and source
Rule
OAC 3701:1-66-02(J)(11) (effective 2024-12-13).
Source
codes.ohio.gov
What it says
(11) Annual evaluation of the integrity of all necessary protective apparel.
This quote was matched word for word to the source page.

Why this matters, in ORSIF's FAQs

Do the rules in Ohio address engineered protection systems used in place of aprons?

UnclearLess certain

Not addressed as such. A whole-body protective barrier is accepted as an alternative to an apron. Under-table drapes do not replace required apparel. The director may grant variances for good cause. Whether an engineered system counts as a whole-body protective barrier is not answered in the text.

Rule text and source
Rule
OAC 3701:1-66-07(A)(11), (B)(1); 3701:1-66-01 (effective 2026-02-16); 3701:1-66-02(A).
Source
codes.ohio.gov
What it says
Such devices will not substitute for wearing obligated protective apparel
This quote was matched word for word to the source page.

About ALARA+

Who operates the equipment

Who may use fluoroscopy equipment and what training the rules ask for.

Who may operate fluoroscopy equipment in Ohio?

In partLess certain

Physicians are exempt from radiologic licensing. Licensed radiographers' scope includes fluoroscopy, under the general supervision of a licensed practitioner. General X-ray machine operators' scope excludes fluoroscopy. In cardiac catheterization, RNs, RCIS staff and certain pre-1995 conditionally licensed operators may only move the patient, move the unit and reset the timer, under in-room personal supervision. PA and NP authority was not settled.

Rule text and source
Rule
ORC 4773.02 (effective 2019-10-17); OAC 3701-72-01, 3701-72-04 (effective 2025-08-31).
Source
codes.ohio.gov
What it says
The following radiologic procedure tasks will be performed only by an Ohio licensed practitioner, radiographer, nuclear medicine technologist, radiation therapist or general x-ray machine operator within his/her scope of practice:
This quote was matched word for word to the source page.

Why this matters, in ORSIF's FAQs

What training or credentials does Ohio require to operate fluoroscopy equipment?

Yes

All fluoroscopy operators, physicians included, and anyone likely to exceed 1 mSv a year from fluoroscopy need at least 2 hours of fluoroscopy-specific radiation protection training. The training must be approved by the facility's radiation expert, be equipment-specific and be documented. 1 hour of retraining is required whenever a person exceeds 30% of the allowable occupational dose in a calendar year. Radiographers need an ODH license (ARRT certification waives the Ohio exam) and 24 continuing education (CE) credits to renew. No separate state fluoroscopy permit was found for physicians.

Rule text and source
Rule
OAC 3701:1-66-07(F)-(G); OAC 3701-72-02.
Source
codes.ohio.gov
What it says
All individuals operating fluoroscopic equipment, and individuals likely to receive an annual effective dose equivalent in excess of one millisievert (one hundred millirem) from participating in fluoroscopic procedures, will receive at least two hours of radiation protection training specific to fluoroscopy
This quote was matched word for word to the source page.

Why this matters, in ORSIF's FAQs

Changes on the way, and recent ones

Rules, bills and guidance the research found from 2025 and 2026, with the ones that touch fluoroscopy teams first.

  • Adopted ruleOAC Chapter 3701-72 radiologic licensing

    Effective (Aug 31, 2025)

    Recently adopted operator licensing rules.

    Source: codes.ohio.gov

  • Adopted ruleOAC 3701:1-66-02, -04 and -07

    Effective (Dec 13, 2024)

    Recently adopted general obligations, QA program and medical fluoroscopic equipment rules.

    Source: codes.ohio.gov

  • BillBill on supervision of general X-ray machine operators by CNPs and PAs

    Bill number and status unconfirmed (Feb 27, 2026)

    An LSC analysis (seen only as a search snippet) describes a bill letting general X-ray machine operators work under the direct supervision of certified nurse practitioners and physician assistants.

    Source: legislature.ohio.gov

  • OtherDraft amendments to OAC Chapters 3701:1-38, 3701:1-66, 3701-72 and 3701-83

    Draft posted for public comment, not yet formally proposed; comments closing October 8, 2026 (Sep 8, 2026)

    ODH posted draft rules for radiation protection standards, radiation-generating equipment, operator licensing and health care facilities. Their contents were not reviewed.

    Source: odh.ohio.gov

  • OtherDraft amendments to OAC Chapter 3701:1-67 (radiation therapy)

    Draft (Nov 24, 2025)

    Draft amendments to the radiation therapy rules, not fluoroscopy.

1 other item the research logged (radiation therapy, industrial machines, fees and other subjects)
  • Adopted ruleOAC 3701:1-66-01 definitions

    Effective (Feb 16, 2026)

    Recently adopted definitions rule. Changes were not compared with the prior version.

    Source: codes.ohio.gov

Also worth knowing

  • Registrants doing fluoroscopically guided interventional (FGI) procedures must have a radiation dose review committee that includes a physician and a technologist who perform the procedures and meets at least annually (3701:1-66-04(D)).
  • Hospital QA must include a quarterly review of occupational exposure records by a certified radiation expert (3701:1-66-04(C)(5)(b)) and policies on pregnant workers (3701:1-66-04(B)(12)).
  • Cumulative air kerma or dose-area product must be recorded for interventional, cardiac, pediatric and pregnant-patient cases (3701:1-66-07(B)(3)).
  • Mobile units and C-arms with SID under 45 cm are exempt from the bucky-slot shield and drape rules. Mini C-arms may be exempt from the apron/barrier rule if a radiation expert documents low dose (3701:1-66-07(D)-(E)).
  • Facilities must have procedures to prevent deceptive exposure of dosimeters (3701:1-38-14(A)(5)).
  • Fluoroscopy training of 2 hours applies to all operators, physicians included.
  • A 2025 PMC review article lists Ohio as requiring 0.50 mm aprons, but the primary text sets 0.25 mm.

What ORSIF could not confirm

Questions the research could not settle. If you know the answer, tell us.

  • What do the 2026-09-08 drafts for 3701:1-38, 3701:1-66 and 3701-72 say, in particular on ALARA, the lens limit, apron rules or engineered systems, and when will they be formally proposed?
  • Does ORC 3748.04 require Ohio's rules to stay compatible with the NRC, so that an NRC ALARA change would bind Ohio?
  • Can physician assistants and nurse practitioners activate fluoroscopy in Ohio? What are the number and status of the bill described in the 2026-02-27 LSC analysis?
  • Would ODH treat an engineered radiation protection system as a whole-body protective barrier, or would it need a variance under 3701:1-66-02(A)?
  • Is there any thyroid or eye protection requirement elsewhere, for example in ODH guidance or the 3701-83 rules?
  • What changed in 3701:1-66-01 (effective 2026-02-16)?

Double-check before you rely on this page

ORSIF compiled this page from the sources linked on it and did its best. It is not comprehensive and it is not always accurate. Rules change, a rule may have been missed or misread, and the sources themselves can be out of date. Treat this page as a starting point.

Where to check

  • The rule itself. Open "Rule text and source" under any answer, or read the sources listed below.
  • The state. Ohio Department of Health. Program contacts in the CRCPD directory.
  • Your radiation safety officer. They know the federal, accreditation and facility requirements that apply at your site, which the state rules do not cover.

What may be missing here

  • 1 of 14 answers on this page are "Not found" or "Unclear". That means ORSIF did not find the answer in the rules it reviewed, not that no rule exists.
  • 2 answers carry a flag or a lower confidence ("Check the source" or "Less certain").
  • Researched November 3, 2026. Rules may have changed since.
  • Only the state's rules for X-ray and fluoroscopy equipment were reviewed.

Found something wrong or missing? Tell ORSIF.

Sources and terms

Terms used on this page
ALARA
As low as reasonably achievable: keep radiation doses as low as is reasonable, below the legal limits.
Agreement State
A state that has signed an agreement with the U.S. Nuclear Regulatory Commission to regulate certain radioactive materials itself.
mSv
Millisievert, a unit of radiation dose. Ten mSv equal one rem, the older unit some rules still use.
Dosimeter
A badge or device worn to measure a worker's radiation dose.
Lead equivalent
The thickness of lead (mm Pb) that would give the same protection as the material described.
NRC
U.S. Nuclear Regulatory Commission
NCRP
National Council on Radiation Protection and Measurements
TEDE
Total effective dose equivalent
FGI
Fluoroscopically guided interventional
NVLAP
National Voluntary Laboratory Accreditation Program
CE
Continuing education
QA
Quality assurance
All 13 sources the research read