Research summary, original research, 2017

Ultra-low radiation dose during electrophysiology procedures using optimized new generation fluoroscopy technology

Crowhurst J, Haqqani H, Wright D, Whitby M, Lee A, Betts J, et al.. Pacing and clinical electrophysiology : PACE. 2017

Read the original paper

In plain language

In one electrophysiology lab, switching to a newer fluoroscopy system set up for low dose cut patient dose area product by about 91% without longer procedures or lower success. Staff dose was not measured directly.

At a glance

Type of study
Original research, Retrospective cohort study
Who was studied
Patients
Summary based on
the abstract

What it found

  • Overall dose area product fell by 91% with the new optimized fluoroscopy system compared with the older system.

    Key figures: 91% (5.0 [2.0-17.0] to 0.45 [0.16-2.61] Gycm2 [P > 0.001])

  • Fluoroscopy times, acquisitions, procedural times and success were largely similar between systems.

  • Dose-area product (DAP) values with the new system were among the lowest published for all procedure groups.

Limits to keep in mind

Single-centre before-after comparison over two consecutive years, so practice changes and case mix may confound results; only patient DAP was measured, not staff dose.

How this summary was made

ORSIF prepared this summary with AI assistance from the abstract of the paper. A program then checked that the passage supporting each finding appears word for word in that source. A person has not yet reviewed each summary, so read the original before relying on a number. ORSIF summarizes research and does not endorse any product or company. Report a problem with this summary.