The Effect Of Low Frame Rate Fluoroscopy On The X-Ray Dose During Coronary Intervention

Sadamatsu K, Nakano Y. Internal medicine (Tokyo, Japan). 2016

Read the original paper

In plain language

In 84 coronary intervention patients at one Japanese centre, fluoroscopy at 7.5 rather than 15 frames per second was associated with about 25% lower air kerma, without longer fluoroscopy time. Staff dose was not measured.

At a glance

Type of study
Original research, Retrospective cohort study
Who was studied
Patients
Size
n = 84 patients
Where
Japan
Summary based on
the abstract

What it found

  • Total air kerma at the interventional reference point was significantly lower with 7.5 frames per second fluoroscopy than with 15 frames per second.

    Key figures: 701.4±427.9 vs. 936.8±623.9 mGy, p=0.02

  • Contrast volume and fluoroscopy time did not differ between frame-rate groups.

  • Authors conclude low frame rate fluoroscopy is safe, feasible and an easy strategy to reduce patient radiation during coronary intervention.

    Key figures: 7.5 frames per second

Limits to keep in mind

Small single-centre retrospective, non-randomised comparison of 84 patients; only patient dose metrics were measured, with no staff dosimetry or clinical outcome data beyond feasibility.

How this summary was made

ORSIF prepared this summary with AI assistance from the abstract of the paper. A program then checked that the passage supporting each finding appears word for word in that source. A person has not yet reviewed each summary, so read the original before relying on a number. ORSIF summarizes research and does not endorse any product or company. Report a problem with this summary.