Research summary, original research, 2012

Reduction of Operator Radiation Dose by a Pelvic Lead Shield During Cardiac Catheterization by Radial Access: Comparison With Femoral Access

Lange HW, von Boetticher H. JACC. Cardiovascular interventions. 2012

Read the original paper

In plain language

In a randomized trial of 210 elective coronary angiograms by one operator, a pelvic lead shield on the patient cut measured operator dose by more than half for radial and femoral access, though radial access still gave higher operator dose.

At a glance

Type of study
Original research, Randomized trial
Who was studied
Interventional cardiologists, Patients
Size
n = 210 patients
Summary based on
the abstract

What it found

  • For radial access, pelvic lead shielding of the patient reduced operator dose by more than half.

    Key figures: 20.9 ± 13.8 μSv to 9.0 ± 5.4 μSv, p < 0.0001

  • For femoral access, pelvic lead shielding reduced operator dose by about 80%.

    Key figures: 15.3 ± 10.4 μSv to 2.9 ± 2.7 μSv, p < 0.0001

  • Dose-area product (DAP)-normalized operator dose fell with shielding by a similar amount for both access routes.

    Key figures: radial 0.94 ± 0.28 to 0.39 ± 0.19; femoral 0.70 ± 0.26 to 0.16 ± 0.13 μSv × Gy(-1) × cm(-2)

  • Even with pelvic shielding, radial access was associated with higher operator dose than femoral access.

Limits to keep in mind

Single operator with one dosimeter outside the apron at chest level, so results may not generalize to other operators, staff or organ doses; abstract only, elective diagnostic angiography, country not stated.

How this summary was made

ORSIF prepared this summary with AI assistance from the abstract of the paper. A program then checked that the passage supporting each finding appears word for word in that source. A person has not yet reviewed each summary, so read the original before relying on a number. ORSIF summarizes research and does not endorse any product or company. Report a problem with this summary.

Prepared November 3, 2026.