Research summary, original research, 2017

Reduction in Radiation Dose in a Pediatric Cardiac Catheterization Lab Using the Philips AlluraClarity X-ray System

Sullivan PM, Harrison D, Badran S, Takao CM, Ing FF. Pediatric cardiology. 2017

Read the original paper

In plain language

In a pediatric cardiac catheterization lab, a newer x-ray system was associated with about 57% lower patient dose area product than an older system, with similar image quality. Staff dose was not measured, though lower patient dose generally lowers scatter.

At a glance

Type of study
Original research, Retrospective cohort study
Who was studied
Patients
Size
n = 762 patients
Summary based on
the abstract

What it found

  • After adjustment for body size, fluoroscopy time and procedure type, the newer AlluraClarity system was associated with a large reduction in patient dose area product across all procedures.

    Key figures: 57.5% (95% confidence interval (CI) 51.5-62.8%, p < 0.001)

  • Median dose-area product (DAP) roughly halved with the new system compared with the older Allura Xper system.

    Key figures: 8661 mGy cm2 vs 4523 mGy cm2 (p < 0.001)

  • Adjusted reductions varied by procedure type, from cardiac biopsies to device occlusions.

    Key figures: 39.0% (95% CI 25.6-50.1%) to 67.6% (95% CI 61.2-72.8%)

  • Image quality scores did not differ significantly between systems.

    Key figures: 4.3 ± 0.8 vs 4.4 ± 0.6, p = 0.62

  • Dose reductions did not significantly differ by patient weight category.

Limits to keep in mind

Single-centre before-after comparison of two non-randomized patient cohorts over different years; only patient DAP was measured, not staff dose.

How this summary was made

ORSIF prepared this summary with AI assistance from the abstract of the paper. A program then checked that the passage supporting each finding appears word for word in that source. A person has not yet reviewed each summary, so read the original before relying on a number. ORSIF summarizes research and does not endorse any product or company. Report a problem with this summary.