Real-time evaluation of operator eye lens radiation dose in interventional radiology: Effects of protective shield position and X-ray beam angle

Sagehashi K, Fujisawa M, Haga Y, Kato T, Takahira S, Sota M, et al.. Radiation Physics and Chemistry. 2026

Read the original paper

In plain language

In cardiovascular procedures on a biplane system, operator eye dose rate rose when the ceiling shield drifted horizontally behind the physician's head, especially in caudal projections; vertical shield height mattered little. Repositioning the shield after angle changes may lower lens dose.

At a glance

Type of study
Original research, Dosimetry measurement study
Who was studied
Interventional cardiologists, Other physicians
Summary based on
the abstract

What it found

  • Horizontal displacement of the ceiling-suspended shield was strongly negatively correlated with operator eye dose rate during fluoroscopy and cine.

    Key figures: R = -0.656 to -0.829, p < 0.01 (fluoroscopy); R = -0.538, p < 0.01 (cine)

  • Eye dose rate increased significantly when the shield was positioned posterior to the physician's head.

  • Dose rate differed by beam angulation, with higher values in caudal projections.

    Key figures: p < 0.001

  • Vertical shield displacement had weak or no correlation with eye dose rate within typical use.

Limits to keep in mind

Abstract only; number of procedures and physicians, country and absolute dose values are not reported. Correlational dosimetry data near the eye, not lens dose under eyewear or health outcomes.

How this summary was made

ORSIF prepared this summary with AI assistance from the abstract of the paper. A program then checked that the passage supporting each finding appears word for word in that source. A person has not yet reviewed each summary, so read the original before relying on a number. ORSIF summarizes research and does not endorse any product or company. Report a problem with this summary.