In cardiovascular procedures on a biplane system, operator eye dose rate rose when the ceiling shield drifted horizontally behind the physician's head, especially in caudal projections; vertical shield height mattered little. Repositioning the shield after angle changes may lower lens dose.
At a glance
- Type of study
- Original research, Dosimetry measurement study
- Who was studied
- Interventional cardiologists, Other physicians
- Topics
- Eye lens and cataractsRoom shielding and protective devicesDosimetry and monitoringDose reduction techniques
- Summary based on
- the abstract
What it found
Horizontal displacement of the ceiling-suspended shield was strongly negatively correlated with operator eye dose rate during fluoroscopy and cine.
Key figures: R = -0.656 to -0.829, p < 0.01 (fluoroscopy); R = -0.538, p < 0.01 (cine)
Eye dose rate increased significantly when the shield was positioned posterior to the physician's head.
Dose rate differed by beam angulation, with higher values in caudal projections.
Key figures: p < 0.001
Vertical shield displacement had weak or no correlation with eye dose rate within typical use.
Limits to keep in mind
Abstract only; number of procedures and physicians, country and absolute dose values are not reported. Correlational dosimetry data near the eye, not lens dose under eyewear or health outcomes.
Related research summaries
- Addressing the Occupational Risk of Radiation Exposure in the Evolving Field of Interventional Echocardiography (2024, Narrative review)
- Radiation Exposure, Training, and Safety in Cardiology (2024, Narrative review)
- Radiation Safety and Protection - StatPearls (2024, Narrative review)
- IU Health Handbook of Fluoroscopy Safety (Guideline or consensus statement)
- Operator Radiation and the Efficacy of Ceiling-Suspended Lead Screen Shielding during Coronary Angiography: An Anthropomorphic Phantom Study Using Real-Time Dosimeters (2017, Phantom or simulation study)
How this summary was made
ORSIF prepared this summary with AI assistance from the abstract of the paper. A program then checked that the passage supporting each finding appears word for word in that source. A person has not yet reviewed each summary, so read the original before relying on a number. ORSIF summarizes research and does not endorse any product or company. Report a problem with this summary.
Prepared November 5, 2026.