In plain language
In a phantom study of coronary angiography, a ceiling-suspended lead screen positioned close to the primary operator, used with table-side shielding, cut measured dose rates by about 94% for the operator and 90% for the assistant, more than other screen positions.
At a glance
- Type of study
- Original research, Phantom or simulation study
- Who was studied
- Phantom or simulation (no people), Interventional cardiologists
- Where
- China
- Topics
- Room shielding and protective devicesDosimetry and monitoringDose reduction techniquesEye lens and cataractsThyroid
- Summary based on
- the full text
What it found
With table-side shielding, placing the ceiling-suspended screen close to the primary operator gave the greatest dose reduction compared with positions near the patient or at the operator's left side.
Key figures: 76.8%, 81.9% and 93.5% reduction for primary operator; p < 0.013 for close-to-PO vs other scenarios
The assistant also received the most protection with the screen close to the primary operator.
Key figures: 70.3%, 76.7% and 90.0% reduction
With the screen close to the primary operator, dose rates at the glabella (eye level) and neck (thyroid level) were reduced by up to about 99%.
Key figures: up to 98.8% (glabella) and 98.6% (neck)
Unshielded primary operator dose rates varied widely by projection and body site, with the lower legs receiving the highest rates; the assistant received significantly less.
Key figures: PO mean 67.2 ± 49.0 μSv/min (range 2.3-227.9); assistant mean 20.1 ± 19.6 μSv/min, P < 0.003
Right anterior oblique (RAO) 30° was the lowest-dose projection for both operators; hypogastrium dose stayed relatively high with some projections even with shielding.
Limits to keep in mind
Single-centre anthropomorphic phantom study with manikin operators in fixed standing positions, one patient body size and short cine acquisitions; dose rates on apron surfaces are not organ doses or health outcomes.
Related research summaries
- Evaluation of the effectiveness of a tunnel-shaped radiation shielding system in CT-guided interventions: Reduction of scattered radiation in phantom experiment (2025, Phantom or simulation study)
- Radiation exposure in interventional cardiology: a narrative review of health effects and novel approaches to ensure radiation safety (2025, Narrative review)
- Radiation Exposure, Training, and Safety in Cardiology (2024, Narrative review)
- A New Design of a Lead-Acrylic Shield for Staff Dose Reduction in Radial and Femoral Access Coronary Catheterization (2015, Phantom or simulation study)
- Radiation Protection in Interventional Radiology (2022, Narrative review)
How this summary was made
ORSIF prepared this summary with AI assistance from the full text of the paper. A program then checked that the passage supporting each finding appears word for word in that source. A person has not yet reviewed each summary, so read the original before relying on a number. ORSIF summarizes research and does not endorse any product or company. Report a problem with this summary.
Prepared November 3, 2026.