Research summary, original research, 2017

Operator Radiation and the Efficacy of Ceiling-Suspended Lead Screen Shielding during Coronary Angiography: An Anthropomorphic Phantom Study Using Real-Time Dosimeters

Jia Q, Chen Z, Jiang X, Zhao Z, Huang M, Li J, et al.. Scientific reports. 2017

Read the original paper

In plain language

In a phantom study of coronary angiography, a ceiling-suspended lead screen positioned close to the primary operator, used with table-side shielding, cut measured dose rates by about 94% for the operator and 90% for the assistant, more than other screen positions.

At a glance

Type of study
Original research, Phantom or simulation study
Who was studied
Phantom or simulation (no people), Interventional cardiologists
Where
China
Summary based on
the full text

What it found

  • With table-side shielding, placing the ceiling-suspended screen close to the primary operator gave the greatest dose reduction compared with positions near the patient or at the operator's left side.

    Key figures: 76.8%, 81.9% and 93.5% reduction for primary operator; p < 0.013 for close-to-PO vs other scenarios

  • The assistant also received the most protection with the screen close to the primary operator.

    Key figures: 70.3%, 76.7% and 90.0% reduction

  • With the screen close to the primary operator, dose rates at the glabella (eye level) and neck (thyroid level) were reduced by up to about 99%.

    Key figures: up to 98.8% (glabella) and 98.6% (neck)

  • Unshielded primary operator dose rates varied widely by projection and body site, with the lower legs receiving the highest rates; the assistant received significantly less.

    Key figures: PO mean 67.2 ± 49.0 μSv/min (range 2.3-227.9); assistant mean 20.1 ± 19.6 μSv/min, P < 0.003

  • Right anterior oblique (RAO) 30° was the lowest-dose projection for both operators; hypogastrium dose stayed relatively high with some projections even with shielding.

Limits to keep in mind

Single-centre anthropomorphic phantom study with manikin operators in fixed standing positions, one patient body size and short cine acquisitions; dose rates on apron surfaces are not organ doses or health outcomes.

How this summary was made

ORSIF prepared this summary with AI assistance from the full text of the paper. A program then checked that the passage supporting each finding appears word for word in that source. A person has not yet reviewed each summary, so read the original before relying on a number. ORSIF summarizes research and does not endorse any product or company. Report a problem with this summary.