In plain language
In phantom-based estimates for paediatric interventional cardiology in Chile, scatter dose to the cardiologist's eyes ranged from 0.20 to 116 µSv per procedure, and doses differed widely between X-ray systems. These are estimates, not health outcomes.
At a glance
- Type of study
- Original research, Phantom or simulation study
- Who was studied
- Interventional cardiologists, Patients, Phantom or simulation (no people)
- Where
- Chile
- Summary based on
- the abstract
What it found
Simulated scatter doses at the cardiologist's eyes varied widely across paediatric procedures.
Key figures: 0.20 to 116 µSv per procedure
Cumulative patient entrance surface air kerma varied widely across simulated procedures.
Key figures: 2 to 1100 mGy
Kerma-area product values ranged widely.
Key figures: 0.30 to 150 Gy cm(2)
Large differences between X-ray systems were found, and the authors call for standardized system settings and protocols in Chile.
Limits to keep in mind
Doses were derived from PMMA phantom measurements combined with average fluoroscopy time and cine frame counts rather than measured on staff during real procedures. The abstract does not state the number of X-ray systems or hospitals.
Related research summaries
- Radiation protection practices in diagnostic and interventional cardiology: Position paper from Portuguese Association of Interventional Cardiology (2025, Guideline or consensus statement)
- Radiation Exposure, Training, and Safety in Cardiology (2024, Narrative review)
- Radiation Safety and Protection - StatPearls (2024, Narrative review)
- Radiation safety in the cath lab: does it still matter? (2020, Editorial or commentary)
- 2018 ACC/HRS/NASCI/SCAI/SCCT Expert Consensus Document on Optimal Use of Ionizing Radiation in Cardiovascular Imaging: Best Practices for Safety and Effectiveness: A Report of the American College of Cardiology Task Force on Expert Consensus Decision Pathways (2019, Guideline or consensus statement)
How this summary was made
ORSIF prepared this summary with AI assistance from the abstract of the paper. A program then checked that the passage supporting each finding appears word for word in that source. A person has not yet reviewed each summary, so read the original before relying on a number. ORSIF summarizes research and does not endorse any product or company. Report a problem with this summary.
Prepared November 5, 2026.