In plain language
Across published dosimetry studies, urologists, orthopedic surgeons, radiologists and gastroenterologists performing fluoroscopy received per-case doses similar to interventional cardiologists, with hand, eye and thyroid doses well above effective dose and wide variation suggesting better shielding and technique could lower exposure.
At a glance
- Type of study
- Systematic review or meta-analysis, Systematic review
- Who was studied
- Interventional radiologists, Urologists, Orthopedic surgeons, Gastroenterologists, Other surgeons, Other physicians, Trainees
- Topics
- Dosimetry and monitoringEye lens and cataractsThyroidBrain tumorsPersonal protective equipmentRoom shielding and protective devicesDose reduction techniquesTraining and educationPatient dose
- Summary based on
- the full text
What it found
Operators performing non-cardiac fluoroscopically guided procedures received per-case effective doses comparable to interventional cardiologists, varying widely by procedure.
Key figures: 1.7-56 μSv percutaneous nephrolithotomy (PCNL); 0.1-101 μSv vertebroplasty; 2.5-88 μSv nailing; 2.0-46 μSv biliary; 2.5-74 μSv transjugular intrahepatic portosystemic shunt (TIPS); 1.8-53 μSv head/neck; 0.2-49 μSv endoscopic retrograde cholangiopancreatography (ERCP)
Doses measured over protective devices at eye, neck and trunk had wide ranges; hand doses were notably higher.
Key figures: eye 19-800 (median 113) μSv; neck 6-1,180 (median 75) μSv; trunk 2-1,600 (median 302) μSv; hand 30-5800 (median 450) μSv
Reported operator doses varied by one to three orders of magnitude between studies for the same procedure, even after normalizing to patient dose, suggesting room for improved protection.
Key figures: 0.02 to 73 μSv/min normalized effective dose
Lens of eye and thyroid doses were much greater than effective dose; brain dose about 5 times lower than lens but still an order of magnitude above effective dose.
Key figures: 1.5 to 1300 μSv per case
Protective measures in included studies substantially reduced dose: leaded screen cut forehead dose 70%, leaded gloves 75%, under-couch shield 8-fold at abdomen, and multiple practice changes cut vertebroplasty whole-body dose from 1440 to 4 μSv.
Key figures: 70%; 75%; 8-fold; 1440 μSv to 4 μSv
Limits to keep in mind
Literature review of small heterogeneous dosimetry studies (2-136 cases each) with inconsistent dosimetry methods and sparse kerma-area product (KAP) data; effective doses were estimated assuming no thyroid shield, and no cumulative or annual doses could be derived.
Related research summaries
- Radiation Exposure, Training, and Safety in Cardiology (2024, Narrative review)
- AST Standards for Ionizing Radiation Exposure (Guideline or consensus statement)
- ICRP Draft - Radiological Protection in Fluoroscopically Guided Procedures (Guideline or consensus statement)
- IU Health Handbook of Fluoroscopy Safety (Guideline or consensus statement)
- Radiation exposure in interventional cardiology: a narrative review of health effects and novel approaches to ensure radiation safety (2025, Narrative review)
How this summary was made
ORSIF prepared this summary with AI assistance from the full text of the paper. A program then checked that the passage supporting each finding appears word for word in that source. A person has not yet reviewed each summary, so read the original before relying on a number. ORSIF summarizes research and does not endorse any product or company. Report a problem with this summary.
Prepared November 3, 2026.