Research summary, systematic review or meta-analysis, 2012

Occupational Radiation Doses to Operators Performing Fluoroscopically-Guided Procedures

Kim KP, Miller DL, Berrington de Gonzalez A, Balter S, Kleinerman RA, Ostroumova E, et al.. Health physics. 2012

Read the original paper

In plain language

Across published dosimetry studies, urologists, orthopedic surgeons, radiologists and gastroenterologists performing fluoroscopy received per-case doses similar to interventional cardiologists, with hand, eye and thyroid doses well above effective dose and wide variation suggesting better shielding and technique could lower exposure.

At a glance

Type of study
Systematic review or meta-analysis, Systematic review
Who was studied
Interventional radiologists, Urologists, Orthopedic surgeons, Gastroenterologists, Other surgeons, Other physicians, Trainees
Summary based on
the full text

What it found

  • Operators performing non-cardiac fluoroscopically guided procedures received per-case effective doses comparable to interventional cardiologists, varying widely by procedure.

    Key figures: 1.7-56 μSv percutaneous nephrolithotomy (PCNL); 0.1-101 μSv vertebroplasty; 2.5-88 μSv nailing; 2.0-46 μSv biliary; 2.5-74 μSv transjugular intrahepatic portosystemic shunt (TIPS); 1.8-53 μSv head/neck; 0.2-49 μSv endoscopic retrograde cholangiopancreatography (ERCP)

  • Doses measured over protective devices at eye, neck and trunk had wide ranges; hand doses were notably higher.

    Key figures: eye 19-800 (median 113) μSv; neck 6-1,180 (median 75) μSv; trunk 2-1,600 (median 302) μSv; hand 30-5800 (median 450) μSv

  • Reported operator doses varied by one to three orders of magnitude between studies for the same procedure, even after normalizing to patient dose, suggesting room for improved protection.

    Key figures: 0.02 to 73 μSv/min normalized effective dose

  • Lens of eye and thyroid doses were much greater than effective dose; brain dose about 5 times lower than lens but still an order of magnitude above effective dose.

    Key figures: 1.5 to 1300 μSv per case

  • Protective measures in included studies substantially reduced dose: leaded screen cut forehead dose 70%, leaded gloves 75%, under-couch shield 8-fold at abdomen, and multiple practice changes cut vertebroplasty whole-body dose from 1440 to 4 μSv.

    Key figures: 70%; 75%; 8-fold; 1440 μSv to 4 μSv

Limits to keep in mind

Literature review of small heterogeneous dosimetry studies (2-136 cases each) with inconsistent dosimetry methods and sparse kerma-area product (KAP) data; effective doses were estimated assuming no thyroid shield, and no cumulative or annual doses could be derived.

How this summary was made

ORSIF prepared this summary with AI assistance from the full text of the paper. A program then checked that the passage supporting each finding appears word for word in that source. A person has not yet reviewed each summary, so read the original before relying on a number. ORSIF summarizes research and does not endorse any product or company. Report a problem with this summary.

Prepared November 3, 2026.