Research summary, original research, 2013

Impact of nonfluoroscopic MediGuide tracking system on radiation exposure in radiofrequency ablation procedures (LESS-RADS registry)-an initial experience

Vallakati A, Reddy YM, Emert M, Janga P, Mansour MC, Heist EK, et al.. Journal of interventional cardiac electrophysiology : an international journal of arrhythmias and pacing. 2013

Read the original paper

In plain language

In 90 electrophysiology procedures, mostly flutter ablations, using the MediGuide catheter tracking system was associated with less than half the fluoroscopy time of conventional mapping. Staff dose was not measured, so the benefit to the electrophysiology (EP) team is inferred.

At a glance

Type of study
Original research, Prospective cohort study
Who was studied
Patients
Size
n = 90 patients (45 MediGuide, 45 matched controls)
Where
United States
Summary based on
the abstract

What it found

  • Fluoroscopy time was substantially lower with the MediGuide nonfluoroscopic tracking system than with conventional mapping systems.

    Key figures: 8 vs. 21 min, p < 0.001

  • Procedure duration was shorter in the MediGuide group.

    Key figures: 103 vs. 142 min, p = 0.03

  • No major complications occurred in either group.

  • MediGuide was associated with more than a 50% reduction in fluoroscopy time.

    Key figures: >50% reduction

Limits to keep in mind

Small observational study of the first 45 cases with matched controls, mostly simple right atrial flutter ablations; only fluoroscopy time was reported, with no staff or patient dose measurements.

How this summary was made

ORSIF prepared this summary with AI assistance from the abstract of the paper. A program then checked that the passage supporting each finding appears word for word in that source. A person has not yet reviewed each summary, so read the original before relying on a number. ORSIF summarizes research and does not endorse any product or company. Report a problem with this summary.

Prepared November 3, 2026.