Research summary, original research, 2017

Image noise reduction technology allows significant reduction of radiation dosage in cardiac device implantation procedures

Hoffmann R, Langenbrink L, Reimann D, Kastrati M, Becker M, Piatkowski M, et al.. Pacing and clinical electrophysiology : PACE. 2017

Read the original paper

In plain language

In 279 cardiac device implantations, a system with image noise reduction cut patient dose-area product by 60% overall, mainly in cardiac resynchronization therapy (CRT) procedures; staff dose was not measured, though lower patient dose generally lowers scatter to the fluoroscopy team.

At a glance

Type of study
Original research, Retrospective cohort study
Who was studied
Patients
Size
n = 279 procedures
Summary based on
the abstract

What it found

  • In CRT implantation, total dose-area product (DAP) was substantially lower with the image noise reduction system than with the conventional C-arm.

    Key figures: 1,544 ± 834 cGycm vs 7,252 ± 6,431 cGycm, P < 0.001

  • Across all device implantation procedures, total DAP fell by 60% with INRT.

    Key figures: 60%

  • In pacemaker/ICD cases, DAP was similar between systems, attributed to pulsed fluoroscopy use on the C-arm and sparse cine.

    Key figures: 423 ± 381 cGycm2 vs 417 ± 517 cGycm

  • CRT reductions came from both lower fluoroscopy DAP and lower cine DAP.

    Key figures: 1,414 ± 757 vs 5,854 ± 6,767 cGycm; 130 ± 106 vs 1,399 ± 1,342 cGycm

  • Patient characteristics and procedure and fluoroscopy times were similar between systems.

Limits to keep in mind

Single-centre, non-randomized before-after comparison of different time periods and different equipment (fixed angiography vs mobile C-arm) with differing fluoroscopy modes; only patient DAP was measured, not staff dose.

How this summary was made

ORSIF prepared this summary with AI assistance from the abstract of the paper. A program then checked that the passage supporting each finding appears word for word in that source. A person has not yet reviewed each summary, so read the original before relying on a number. ORSIF summarizes research and does not endorse any product or company. Report a problem with this summary.