In plain language
In a phantom endoscopic retrograde cholangiopancreatography (ERCP) setup, a couch-mounted upright shield cut scattered radiation at operator eye level by 90% and at a seated anesthesiologist's eye level by 86%, provided it was tall enough. Measured staff eye doses in real procedures were not studied.
At a glance
- Type of study
- Original research, Phantom or simulation study
- Who was studied
- Phantom or simulation (no people), Gastroenterologists, Anesthesia staff
- Where
- Japan
- Topics
- Room shielding and protective devicesEye lens and cataractsDosimetry and monitoringThyroidBrain tumors
- Summary based on
- the full text
What it found
A 40 cm upright shield on the C-arm reduced scattered radiation at the operator's eye level by 90%.
Key figures: 90%
Shield height was critical: raising it from 30 to 40 cm raised eye-level protection from 33% to 90%, attributed to closing the gap to the image detector.
Key figures: 33% to 90%
The head-side shield gave 86% protection at 100 cm (seated anesthesiologist eye level) but under 10% at 150 cm.
Key figures: 86% at 100 cm; <10% at 150 cm
Without shielding, operator eye-level scatter per kerma-area product (PKA) on the C-arm was 4.7 times that of an over-couch system with an ERCP-specific curtain; the 40 cm shield brought it below the curtain system.
Key figures: 4.7 times
The table-suspended under-curtain alone gave minimal eye-level shielding; the 20 cm shield gave 17% at 150 cm but 98% at 100 cm.
Key figures: 17% at 150 cm; 98% at 100 cm
Limits to keep in mind
Phantom study with a survey meter measuring H*(10) rather than Hp(3) eye dose, fixed staff positions, single setup; the shield does not cover nurses and clinical staff movement was not assessed.
Related research summaries
- Occupational Radiation Doses to Operators Performing Fluoroscopically-Guided Procedures (2012, Systematic review)
- Radiation exposure in interventional cardiology: a narrative review of health effects and novel approaches to ensure radiation safety (2025, Narrative review)
- Radiation Exposure, Training, and Safety in Cardiology (2024, Narrative review)
- Occupational Risks of Chronic Low Dose Radiation Exposure in Cardiac Catheterisation Laboratory (2013, Narrative review)
- Evaluation of the effectiveness of a tunnel-shaped radiation shielding system in CT-guided interventions: Reduction of scattered radiation in phantom experiment (2025, Phantom or simulation study)
How this summary was made
ORSIF prepared this summary with AI assistance from the full text of the paper. A program then checked that the passage supporting each finding appears word for word in that source. A person has not yet reviewed each summary, so read the original before relying on a number. ORSIF summarizes research and does not endorse any product or company. Report a problem with this summary.
Prepared November 3, 2026.