Research summary, original research, 2019

From near-zero to zero fluoroscopy catheter ablation procedures

Santoro A, Di Clemente F, Baiocchi C, Zacà V, Bianchi C, Bellini C, et al.. Journal of cardiovascular electrophysiology. 2019

Read the original paper

In plain language

In 525 catheter ablations at one centre, electroanatomical mapping allowed 96% of later procedures to be done with no fluoroscopy, without loss of success or added complications. Staff dose was not measured, but eliminating fluoroscopy removes the radiation source.

At a glance

Type of study
Original research, Retrospective cohort study
Who was studied
Patients
Size
n = 525 procedures
Summary based on
the abstract

What it found

  • Overall, most catheter ablations in the series were performed with no fluoroscopy at all.

    Key figures: 78.5% of 525 procedures

  • Zero-fluoroscopy procedures rose from about a third in the early period to nearly all in the later period, consistent with a learning curve.

    Key figures: 38.5% (2011-2013) vs 96.2% (2014-2017)

  • Fluoroscopy time fell sharply in the later group.

    Key figures: 1.4 ± 19.6 s vs 556.92 ± 520.76 s; P < .001

  • Eliminating fluoroscopy did not change acute success, complications, or procedure duration.

Limits to keep in mind

Single-centre observational before-after comparison reported only as an abstract; procedures needing transseptal puncture and nonidiopathic ventricular tachycardia (VT) were excluded, and staff dose was not measured directly.

How this summary was made

ORSIF prepared this summary with AI assistance from the abstract of the paper. A program then checked that the passage supporting each finding appears word for word in that source. A person has not yet reviewed each summary, so read the original before relying on a number. ORSIF summarizes research and does not endorse any product or company. Report a problem with this summary.

Prepared November 3, 2026.