In plain language
Measured scatter around computed tomography (CT) scanners can be estimated from the dose-length product (0.36 for body, 0.14 for head scans, µGy per mGy cm at 1 m), helping plan shielding and assess dose to staff present during CT fluoroscopy.
At a glance
- Type of study
- Original research, Dosimetry measurement study
- Who was studied
- Phantom or simulation (no people)
- Where
- United Kingdom
- Summary based on
- the abstract
What it found
Recommended scatter factors to estimate scatter air kerma at 1 m from the CT isocentre differ for body and head scans.
Key figures: 0.36 µGy (mGy cm)(-1) body; 0.14 µGy (mGy cm)(-1) head
CT gantries transmit only a small fraction of scatter, which can be accounted for in shielding design.
Key figures: 10%
The factors can support risk assessments for staff who must be present during CT fluoroscopy procedures.
Limits to keep in mind
Abstract only; number of scanners and measurement conditions not stated, and results are physical measurements of scatter rather than staff doses or health outcomes.
Related research summaries
- Technical Note: Assessment Of Scatter Originating From The X-Ray Tube Collimator Assembly Of Modern Angiography Systems (2019, Phantom or simulation study)
- Radiation protection: safety measures and knowledge among interventional radiologists - UK-based analysis (2025, Cross-sectional survey)
- Accidents of occupational exposure to ionizing radiation in radiology departments: risks, prevention, and management (2026, Narrative review)
- Occupational Protection in Interventional Radiology. A Joint Guideline of the Cardiovascular and Interventional Radiological Society of Europe and the Society of Interventional Radiology (2026, Guideline or consensus statement)
- SCAI/ASE/HRS/SIR/SVS Expert Consensus Statement on Enhanced Radiation Protection: Time for Mandatory and Urgent Action (2026, Guideline or consensus statement)
How this summary was made
ORSIF prepared this summary with AI assistance from the abstract of the paper. A program then checked that the passage supporting each finding appears word for word in that source. A person has not yet reviewed each summary, so read the original before relying on a number. ORSIF summarizes research and does not endorse any product or company. Report a problem with this summary.
Prepared November 5, 2026.