Technical Note: Assessment Of Scatter Originating From The X-Ray Tube Collimator Assembly Of Modern Angiography Systems

Tao AT, Miller D, Hindal M, Fetterly KA. Medical physics. 2019

Read the original paper

In plain language

In phantom measurements on clinical angiography systems, scatter from the x-ray tube collimator housing made up 20-50% of total scatter in PA projection. Adding 0.5 mm lead inside the cover lowered its share from 28% to 16%. Health effects on the fluoroscopy team were not measured.

At a glance

Type of study
Original research, Phantom or simulation study
Who was studied
Phantom or simulation (no people)
Summary based on
the abstract

What it found

  • In the PA projection, scatter from the x-ray tube collimator assembly made up a substantial share of total scatter across systems.

    Key figures: 20-50% of total scatter

  • The collimator contribution varied with projection angle.

    Key figures: 5% to 56% for -90º to 90º projections

  • Adding 0.5 mm lead inside the collimator cover reduced the collimator share of scatter.

    Key figures: 28% to 16% of total scatter

  • Systems with kerma-area product meters showed more collimator scatter.

  • Authors suggest radiation safety practices and guidelines be revised to account for collimator scatter.

Limits to keep in mind

Phantom-based measurements on a limited, unspecified number of angiography systems, with lead modification tested on only two; abstract only, so system count and absolute dose values are not reported.

How this summary was made

ORSIF prepared this summary with AI assistance from the abstract of the paper. A program then checked that the passage supporting each finding appears word for word in that source. A person has not yet reviewed each summary, so read the original before relying on a number. ORSIF summarizes research and does not endorse any product or company. Report a problem with this summary.