Effects of reducing frame rate from 7.5 to 4 frames per second on radiation exposure in transcatheter atrial septal defect closure

Boudjemline Y. Cardiology in the young. 2018

Read the original paper

In plain language

In 134 transcatheter atrial septal defect closures, using 4 instead of 7.5 fluoroscopy frames per second cut patient air kerma and dose area product by about two thirds with similar success. Staff dose was not measured but would likely fall too.

At a glance

Type of study
Original research, Retrospective cohort study
Who was studied
Patients
Size
n = 134 procedures
Summary based on
the abstract

What it found

  • Median total air kerma was lower with 4 frames per second than with 7.5 frames per second.

    Key figures: 4 versus 1.3 mGy [p=0.00012]

  • Median dose area product was lower with 4 frames per second.

    Key figures: 43.7 versus 13.1 μGy.m2

  • Procedural success was similar between frame rates.

    Key figures: 100 versus 98.8%

  • Authors recommend adopting the lower frame rate to reduce exposure to patients and laboratory personnel.

Limits to keep in mind

Single-centre retrospective comparison of 134 procedures using patient dose metrics only; staff exposure was not measured and the country and patient ages are not stated in the abstract.

How this summary was made

ORSIF prepared this summary with AI assistance from the abstract of the paper. A program then checked that the passage supporting each finding appears word for word in that source. A person has not yet reviewed each summary, so read the original before relying on a number. ORSIF summarizes research and does not endorse any product or company. Report a problem with this summary.