In plain language
In procedures by two interventional radiologists, adding real-time staff dose monitoring to patient dose monitoring was followed by lower patient dose (mean kerma-area product (KAP) 47 to 37 Gy·cm2). Operator and patient doses correlated strongly; this before-after design cannot prove cause.
At a glance
- Type of study
- Original research, Prospective cohort study
- Who was studied
- Interventional radiologists, Patients
- Summary based on
- the abstract
What it found
Total patient KAP fell when real-time occupational dose monitoring was added to patient dose monitoring, for both operators combined.
Key figures: period 1, 47 Gy ∙ cm(2) ± 67/41 Gy ∙ cm(2); period 2, 37 Gy ∙ cm(2) ± 69/34 Gy ∙ cm(2); P < .05
Patient dose and occupational dose were strongly positively correlated.
Key figures: r = 0.88
KAP declined in most procedure types during the combined-monitoring period.
Key figures: 15 of 19 types of fluoroscopically guided interventionals (FGIs)
Mean operator dose per intervention was low with real-time monitoring.
Key figures: 4.6 µSv mean accumulated dose per intervention; 0.24 mSv/h mean dose rate
The KAP reduction was also seen for each operator individually.
Key figures: P < .05
Limits to keep in mind
Before-after comparison involving only two experienced interventional radiologists, without randomization or control for other changes over time; the number of procedures is not stated and operator dose reduction is inferred from correlation rather than a between-period comparison.
Related research summaries
- Reduction of Radiation Exposure Using Dynamic Trace Digital Angiography and Spot Fluoroscopy During Adrenal Venous Sampling (2017, Original research)
- Radiation Protection in Interventional Radiology (2022, Narrative review)
- AAPM Medical Physics Practice Guideline 12.a - Fluoroscopy Dose Management (2022, Guideline or consensus statement)
- IU Health Handbook of Fluoroscopy Safety (Guideline or consensus statement)
- Fluoroscopy-minimization strategy for catheter ablation of Supraventricular Tachycardia by wOmen OPerators: the STOOP Multicentre Registry (2026, Prospective cohort study)
How this summary was made
ORSIF prepared this summary with AI assistance from the abstract of the paper. A program then checked that the passage supporting each finding appears word for word in that source. A person has not yet reviewed each summary, so read the original before relying on a number. ORSIF summarizes research and does not endorse any product or company. Report a problem with this summary.
Prepared November 3, 2026.