Please double-check this information. This is an ORSIF research summary, researched November 3, 2026. It is not comprehensive and not always accurate: rules change, and answers may be incomplete, out of date or wrong. It is education, not legal advice. Not yet reviewed by a regulatory expert. Confirm each requirement at the source linked beside the answer, with the state program and with your facility's radiation safety officer. How this was made

In brief

ORSIF's research identified the Colorado Department of Public Health and Environment (CDPHE), Hazardous Materials and Waste Management Division, Radiation Control Program; rules adopted by the State Board of Health as the agency that regulates X-ray and fluoroscopy equipment in Colorado. Colorado is a U.S. Nuclear Regulatory Commission (NRC) Agreement State: it has signed an agreement with the NRC to regulate certain radioactive materials itself.

Colorado's ALARA requirement is in its unified Part 4 (§4.5.2), which covers both radioactive-materials licensees and X-ray registrants; Part 6 points back to it. Colorado uses the standard federal dose limits (50 mSv whole body, 150 mSv lens, 500 mSv skin/extremity, 5 mSv embryo/fetus) and has not adopted a lower lens limit. Fluoroscopy staff must wear aprons or stand behind shields of at least 0.25 mm lead equivalent, and aprons must be checked visually and by touch every year. Barriers and shields are accepted as alternatives, but engineered protection systems are not named. Since 2021, fluoroscopy operators must be physicians, certain American Registry of Radiologic Technologists (ARRT) or radiologist-assistant credential holders, or registered fluoroscopy operators (PAs, APRNs and certified cardiovascular lab specialists with 40 hours of coursework and 40 hours of supervised clinical experience). The Part 6 findings are based on the 2020 text and need checking against the 2024 version.

The rules, question by question

Every answer names the rule it comes from. Open "Rule text and source" to read the citation, the source page and, where it could be matched, the exact wording.

Where the rules live

Which rules apply, and whether ALARA is among them.

Is Colorado an NRC Agreement State?

Yes

Yes. Colorado is not on the NRC list of non-Agreement States, and its rules include a radioactive-materials licensing part.

Rule text and source
Rule
NRC list of Agreement States, updated October 29, 2026; 6 CCR 1007-1 Part 03
Source
nrc.gov
coloradosos.gov
What it says
RADIATION CONTROL - LICENSING OF RADIOACTIVE MATERIAL
This quote was matched word for word to the source page.

Why this matters, in ORSIF's FAQs

Which rules cover X-ray and fluoroscopy equipment in Colorado?

Yes

6 CCR 1007-1: Part 4 (Standards for Protection Against Radiation, eff. 12/15/2023), Part 6 (X-Ray Imaging in the Healing Arts, eff. 04/14/2024), Part 2 (Registration of Radiation Machines, Facilities and Services, eff. 04/14/2024), and Part 1 (General Provisions, eff. 11/14/2017). The Part 4 text was read in a copy filed with the NRC, not the official SOS PDF.

Rule text and source
Rule
6 CCR 1007-1 Part 4, §4.1.3.1
Source
coloradosos.gov
What it says
This Part 4 establishes standards for protection against ionizing radiation resulting from activities conducted pursuant to licenses or registrations issued by the Department.
This quote was matched word for word to the source page.

Why this matters, in ORSIF's FAQs

Does Colorado require X-ray users to keep doses as low as reasonably achievable (ALARA)?

Yes

Yes. §4.5.2 requires every licensee or registrant, including X-ray registrants, to use procedures and engineering controls that keep occupational doses ALARA.

Where ALARA lives in Colorado: Only in a rule shared with radioactive materials. The only operative ALARA sentence is in the unified Part 4 (§4.5.2), which applies to both licensees and registrants (§4.1.4.1). In the 2020 text, Part 6 cross-references Part 4 (§6.3.3.1) rather than creating its own ALARA duty, apart from one X-ray-specific line on auxiliary equipment designed to minimize exposure. (§4.1.4.1; §4.5.2; §6.3.3.1 (2020 version))

Rule text and source
Rule
6 CCR 1007-1 §4.5.2 (eff. 12/15/2023 version)
Source
nrc.gov
What it says
The licensee or registrant shall use, to the extent practical, procedures and engineering controls based upon sound radiation protection principles to achieve occupational doses and doses to members of the public that are as low as is reasonably achievable (ALARA).
This quote was matched word for word to the source page.

Why this matters, in ORSIF's FAQs

Dose limits

The annual limits for workers, written in millisieverts (mSv). Ten mSv equal one rem.

What is the annual eye-lens dose limit in Colorado?

Yes

150 mSv (15 rem) per year. Colorado has not adopted a lower lens limit.

Rule text and source
Rule
§4.6.1.2
Source
nrc.gov
What it says
(1) A lens dose equivalent of 0.15 Sv (15 rem), and (2) A shallow dose equivalent of 0.5 Sv (50 rem) to the skin of the whole body or to the skin of any extremity.
This quote was matched word for word to the source page.

Why this matters, in ORSIF's FAQs

What is the annual limit for the skin, hands and feet in Colorado?

Yes

500 mSv (50 rem) per year shallow dose to the skin of the whole body or any extremity.

Rule text and source
Rule
§4.6.1.2
Source
nrc.gov
What it says
(1) A lens dose equivalent of 0.15 Sv (15 rem), and (2) A shallow dose equivalent of 0.5 Sv (50 rem) to the skin of the whole body or to the skin of any extremity.
This quote was matched word for word to the source page.

Why this matters, in ORSIF's FAQs

What limit applies to a declared pregnant worker in Colorado?

Yes

5 mSv (0.5 rem) to the embryo/fetus over the whole pregnancy, with efforts to avoid substantial month-to-month variation in dose (§4.13.2).

Rule text and source
Rule
§4.13.1
Source
nrc.gov

Why this matters, in ORSIF's FAQs

Dosimeters

When a worker must wear a dosimeter, and where.

When must a worker in Colorado wear a dosimeter?

Yes

Monitoring is required for adults likely to receive more than 10% of the §4.6.1 limits; minors likely to exceed 1 mSv deep dose, 1.5 mSv lens dose or 5 mSv skin/extremity dose; declared pregnant women likely to exceed 1 mSv deep dose over the pregnancy; and anyone entering high or very high radiation areas. In the 2020 text, §6.3.3.11 applies these duties to X-ray staff.

Rule text and source
Rule
§4.18.1.1-4.18.1.4
Source
nrc.gov
What it says
Adults likely to receive, in 1 year from sources external to the body, a dose in excess of 10 percent of the limits in 4.6.1;
This quote was matched word for word to the source page.

Why this matters, in ORSIF's FAQs

Where must the dosimeter be worn in Colorado when a protective apron is worn?

Yes

Under 145 kVp with lead garments, the registrant may calculate assigned dose using NRC RIS 2002-06 methods, Department-approved methods, or Colorado's equations: 0.06 × (collar - waist) + waist without a thyroid collar, and 0.02 × (collar - waist) + waist with one. Without an approved method, deep-dose equivalent replaces effective dose equivalent. National Council on Radiation Protection and Measurements (NCRP) 122 and Webster are not named. No explicit badge-location rule was found; assigned deep dose must be for the part of the body receiving the highest exposure.

Rule text and source
Rule
§4.6.3.1, 4.6.3.2, 4.6.3.5; §6.3.3.11 (2020)
Source
nrc.gov
What it says
In the case of occupational exposures to x-rays with accelerating voltages of less than 145 kVp and where the worker utilizes lead garment protection, the registrant may calculate the assigned dose equivalent using methods discussed in NRC Regulatory Information Summary (RIS) 2002-06
This quote was matched word for word to the source page.

Why this matters, in ORSIF's FAQs

Aprons and alternatives

Protective apparel, how it is checked, and whether other systems are addressed.

Must staff in Colorado fluoroscopy rooms wear protective aprons or other apparel?

YesLess certain

Based on the 2020 text: ancillary shielding must be available and used for stationary fluoroscopes. Where sterile fields prevent normal barriers, everyone except the patient must wear aprons or stand behind a shield of at least 0.25 mm lead equivalent. Staff must be protected by at least 0.25 mm lead equivalent from scatter and 0.5 mm from the direct beam, with a radiation safety officer (RSO) exception. No thyroid shield or eyewear requirement was found for human fluoroscopy staff.

Rule text and source
Rule
§6.5.11.1-6.5.11.2; §6.3.3.5 (2020 version)
Source
sos.state.co.us
What it says
All persons, except the patient, in the room where fluoroscopy is performed shall wear protective apparel (aprons) or shall be positioned behind a stationary or portable shield that provides a lead equivalent shielding of at least 0.25mm;
This quote was matched word for word to the source page.

Why this matters, in ORSIF's FAQs

Does Colorado require protective aprons to be inspected?

YesLess certain

Yes, annually under a written procedure, by visual inspection plus a tactile test; X-ray or fluoroscopic imaging of aprons is not required. Garments must be labeled with lead equivalence and hung, failed items must be removed from service, repaired or replaced, and records kept for 3 years (2020 text).

Rule text and source
Rule
§6.3.3.6(1)-(4) (2020 version)
Source
sos.state.co.us
What it says
All protective apparel and auxiliary shields shall be evaluated annually for integrity.
This quote was matched word for word to the source page.

Why this matters, in ORSIF's FAQs

Do the rules in Colorado address engineered protection systems used in place of aprons?

In partLess certain

Engineered radiation protection systems are not addressed by name. Whole body protective barriers and stationary or portable shields are accepted as alternatives to garments. It is unclear whether ceiling- or floor-mounted systems qualify, and no approval process was found.

Rule text and source
Rule
§6.3.3.5(2); §6.5.11.2(2) (2020 version)
Source
sos.state.co.us
What it says
All persons shall be protected from scatter radiation by protective garments, safety equipment or whole body protective barriers of at least 0.25 millimeter lead equivalent material
This quote was matched word for word to the source page.

About ALARA+

Who operates the equipment

Who may use fluoroscopy equipment and what training the rules ask for.

Who may operate fluoroscopy equipment in Colorado?

Yes

Since January 1, 2021, fluoroscopy operators on living humans must be Colorado-licensed physicians, RRA/RPA meeting Appendix 2G, technologists with R.T.(R), (CV), (CI), (VI) or (T), or registered fluoroscopy operators. Registration is limited to licensed PAs, licensed APRNs and nationally certified cardiovascular lab specialists aged 18 or older. Limited-scope operators may not do fluoroscopy, and no category was found for non-APRN RNs. The 2020 text requires direct supervision; the current §6.3.1.6 adds a personal-supervision route for operators who meet Appendix 2O training.

Rule text and source
Rule
§2.4.5.5(1); Appendix 2O; §6.5.12.1 (2020); §6.3.1.6
Source
sos.state.co.us
What it says
On or after January 1, 2021, each individual operating a fluoroscopy imaging system on living humans shall be registered with the department as a fluoroscopy operator consistent with 2.4.5.5(2) or 2.4.5.5(3), except for:
This quote was matched word for word to the source page.

Why this matters, in ORSIF's FAQs

What training or credentials does Colorado require to operate fluoroscopy equipment?

Yes

Registration requires at least 40 hours of coursework, 40 hours of clinical experience under the personal supervision of a Colorado-licensed physician, and a credential: a passing ARRT fluoroscopy exam for PAs and APRNs, or CCI RCIS/RCES or another Department-approved credential for others. Registration lasts two years and is renewed by resubmitting a current license or certification; no continuing education (CE) hours were found. Physicians are deemed trained, but everyone who operates or supervises fluoroscopy needs documented training under §2.6.1.5, which since 2022 must cover patient and staff radiation protection, dose units and the outputs of each operating mode.

Rule text and source
Rule
Appendix 2O (2O.1-2O.2); §2.4.5.5(3)(e); §2.6.1.5
Source
sos.state.co.us
What it says
Has completed a course that includes at least forty (40) hours of education on topics that include, but are not limited to, radiation physics, radiation biology, radiation safety and radiation management applicable to fluoroscopy;
This quote was matched word for word to the source page.

Why this matters, in ORSIF's FAQs

Also worth knowing

  • Since 2022, facilities performing fluoroscopically guided interventional (FGI) procedures must have an FGI Procedure Committee that includes a supervising physician, a registered medical physicist and, where applicable, a technologist, meets at least every 12 months, and focuses its dose management on patients (§6.5.15, 2020 version).
  • Registrants must keep a list of fluoroscopy operators and supervisors, updated at least once a year, as part of the Part 4 radiation protection program (§2.6.1.1(2)).
  • Part 6 answers are based on the superseded 10/15/2020 text; the current version took effect 04/14/2024.
  • §4.6.3.1 refers to effective-dose methods approved by the NRC.
  • Thyroid shield and eye protection appear only in the veterinary rules (§6.8.2.1(4)).

What ORSIF could not confirm

Questions the research could not settle. If you know the answer, tell us.

  • Read the official 04/14/2024 Part 6 PDF to confirm the wording and numbering of §6.3.3.5, §6.3.3.6, §6.3.3.11, §6.5.11, §6.5.12.1 and §6.5.15, and check eDocket 2024-00016 and any 2021-2022 dockets.
  • Confirm which operators need direct supervision (§6.5.12.1, 2020) versus personal supervision (§6.3.1.6(4), 2024), including Part 1's definition of personal supervision.
  • Pending 2025-2026 rules or bills were not searched (SOS eDocket, Colorado Register, CDPHE rulemaking pages, leg.colorado.gov); also ask how CDPHE will handle any NRC compatibility change to 10 CFR 20.1101.
  • Whether CDPHE treats ceiling- or table-mounted shielding systems as whole body protective barriers (see cdphe.colorado.gov/hm/xray-guidance-and-policy or ask the program).
  • Which methods NRC RIS 2002-06 covers (e.g., Webster, NCRP 122), and whether CDPHE has approved others under §4.6.3.5.
  • Confirm the agency website by opening cdphe.colorado.gov/hm/radiation-management.

Double-check before you rely on this page

ORSIF compiled this page from the sources linked on it and did its best. It is not comprehensive and it is not always accurate. Rules change, a rule may have been missed or misread, and the sources themselves can be out of date. Treat this page as a starting point.

Where to check

What may be missing here

  • 0 of 14 answers on this page are "Not found" or "Unclear". That means ORSIF did not find the answer in the rules it reviewed, not that no rule exists.
  • 3 answers carry a flag or a lower confidence ("Check the source" or "Less certain").
  • Researched November 3, 2026. Rules may have changed since.
  • Only the state's rules for X-ray and fluoroscopy equipment were reviewed.

Found something wrong or missing? Tell ORSIF.

Sources and terms

Terms used on this page
ALARA
As low as reasonably achievable: keep radiation doses as low as is reasonable, below the legal limits.
Agreement State
A state that has signed an agreement with the U.S. Nuclear Regulatory Commission to regulate certain radioactive materials itself.
mSv
Millisievert, a unit of radiation dose. Ten mSv equal one rem, the older unit some rules still use.
Dosimeter
A badge or device worn to measure a worker's radiation dose.
Lead equivalent
The thickness of lead (mm Pb) that would give the same protection as the material described.
NRC
U.S. Nuclear Regulatory Commission
NCRP
National Council on Radiation Protection and Measurements
TEDE
Total effective dose equivalent
ARRT
American Registry of Radiologic Technologists
FGI
Fluoroscopically guided interventional
RSO
Radiation safety officer
CE
Continuing education
All 9 sources the research read

Found something out of date or wrong? Tell ORSIF. Corrections are checked against the rule text and the page is updated.