Please double-check this information. This is an ORSIF research summary, researched November 3, 2026. It is not comprehensive and not always accurate: rules change, and answers may be incomplete, out of date or wrong. It is education, not legal advice. Not yet reviewed by a regulatory expert. The research for this state ended before every question was fully answered, so some answers may be incomplete. Confirm each requirement at the source linked beside the answer, with the state program and with your facility's radiation safety officer. How this was made

In brief

ORSIF's research identified the Radiological Health Program, State Public Health Laboratories (Alaska Department of Health, Division of Public Health, Section of Laboratories) as the agency that regulates X-ray and fluoroscopy equipment in Alaska. Alaska is not a U.S. Nuclear Regulatory Commission (NRC) Agreement State, so the NRC regulates radioactive materials there directly. The state still regulates X-ray machines.

Alaska's fluoroscopy rules are in 7 AAC 18, which covers all radiation sources and has a chapter-wide ALARA requirement (18.105(c)). Alaska is not an NRC Agreement State, so an NRC rule requiring Agreement States to remove ALARA would not bind it directly. Worker dose limits are 5 rem (50 mSv) whole body and 15 rem (150 mSv) for the lens of the eye; no lower eye limit was found. Anyone in the fluoroscopy room likely to get 5 mR per hour or more must wear an apron and gloves of at least 0.25 mm lead equivalent, and protective drapes and panels may not substitute for an apron. No rule on engineered shielding systems used instead of aprons was found. No operator license or permit appeared in the rules read; the facility must ensure operators are trained and competent, and each fluoroscopy operator needs at least 10 hours of instruction. The rule text was read only on secondary sources (Cornell LII, Justia), and the research for Alaska ended before every question was fully checked.

The rules, question by question

Every answer names the rule it comes from. Open "Rule text and source" to read the citation, the source page and, where it could be matched, the exact wording.

Where the rules live

Which rules apply, and whether ALARA is among them.

Is Alaska an NRC Agreement State?

No

No. Alaska is not an NRC Agreement State. This comes from the NRC list (updated 2026-10-29). The program page says it works with the NRC.

Rule text and source
Rule
NRC list of Agreement States, updated October 29, 2026; program webpage.
Source
nrc.gov
health.alaska.gov

Why this matters, in ORSIF's FAQs

Which rules cover X-ray and fluoroscopy equipment in Alaska?

YesLess certain

7 AAC 18, Radiation Sources and Radiation Protection. Article 2 covers applicability and registration. Article 3 holds the general protection standards. Article 4 covers healing arts uses, including 7 AAC 18.440 on fluoroscopic installations. Article 6 holds definitions. 7 AAC 19 covers dental X-ray only. The research read the rule text only on secondary sources (Cornell LII, Justia).

Rule text and source
Rule
7 AAC 18.105(a); 7 AAC 18, Articles 3 and 4.
Source
regulations.justia.com
What it says
the provisions of 7 AAC 18.110 - 7 AAC 18.990 apply to a person in this state who receives, possesses, uses, transfers, owns, or acquires a radiation source.
This quote was matched word for word to the source page.

Why this matters, in ORSIF's FAQs

Does Alaska require X-ray users to keep doses as low as reasonably achievable (ALARA)?

Yes

Yes. 7 AAC 18.105(c) requires registrants and operators to apply the principles of ALARA, and 18.105(d) defines it. Operator training must also cover ALARA (18.420(b)(1)(B)).

Where ALARA lives in Alaska: In a shared rule and in X-ray-specific language. The main requirement, 7 AAC 18.105(c), is chapter-wide and applies to all radiation sources; the chapter uses 'registrant' for both X-ray and radioactive material. The X-ray operator training curriculum (18.420(b)(1)(B)) also requires teaching ALARA. Because Alaska is not an Agreement State, an NRC rule requiring Agreement States to drop ALARA would not legally reach this chapter. (7 AAC 18.105(c); 7 AAC 18.150; 7 AAC 18.420(b)(1)(B))

Rule text and source
Rule
7 AAC 18.105(c)-(d); 7 AAC 18.420(b)(1)(B)
Source
regulations.justia.com
What it says
registrants and operators shall apply the principles of ALARA to ensure that registrants, operators, and the general public are exposed only to the lowest amount of ionizing radiation necessary to accomplish an intended purpose.
This quote was matched word for word to the source page.

Why this matters, in ORSIF's FAQs

Dose limits

The annual limits for workers, written in millisieverts (mSv). Ten mSv equal one rem.

What is the annual whole-body dose limit in Alaska?

Yes

5 rem (50 mSv) total effective dose equivalent per year, or 50 rem (500 mSv) to any single organ other than the lens, whichever is lower. Doses from multiple registrants are combined.

Rule text and source
Rule
7 AAC 18.210(a)-(c)
Source
regulations.justia.com

Why this matters, in ORSIF's FAQs

What is the annual eye-lens dose limit in Alaska?

Yes

15 rem (150 mSv) per year. No lower lens limit has been adopted; a circa-2013 proposal to cut it to 5 rem appears not to have been adopted.

Rule text and source
Rule
7 AAC 18.210(a)
Source
regulations.justia.com
What it says
(A) lens dose equivalent of 15 rems (0.15 Sv); and (B) shallow-dose equivalent of 50 rem (0.5 Sv) to the skin of the whole body or to the skin of any extremity.
This quote was matched word for word to the source page.

Why this matters, in ORSIF's FAQs

What is the annual limit for the skin, hands and feet in Alaska?

Yes

Shallow-dose equivalent of 50 rem (500 mSv) per year to the skin of the whole body or any extremity.

Rule text and source
Rule
7 AAC 18.210(a)
Source
regulations.justia.com
What it says
(A) lens dose equivalent of 15 rems (0.15 Sv); and (B) shallow-dose equivalent of 50 rem (0.5 Sv) to the skin of the whole body or to the skin of any extremity.
This quote was matched word for word to the source page.

Why this matters, in ORSIF's FAQs

What limit applies to a declared pregnant worker in Alaska?

Not foundCheck the source

not found

Rule text and source
Rule
7 AAC 18.255 (not read)

Why this matters, in ORSIF's FAQs

Dosimeters

When a worker must wear a dosimeter, and where.

When must a worker in Alaska wear a dosimeter?

YesCheck the source

An individual dosimeter is required for anyone (other than a patient) in a controlled area who receives, or is likely to receive, more than 10% of an annual limit. The threshold is 5% for minors. Anyone entering a high or very high radiation area must also be monitored.

Rule text and source
Rule
7 AAC 18.280(a)(1)-(3)
Source
regulations.justia.com

Why this matters, in ORSIF's FAQs

Where must the dosimeter be worn in Alaska when a protective apron is worn?

Not foundLess certain

not found

Rule text and source
Rule
7 AAC 18.210(b); 7 AAC 18.280(a)
Source
regulations.justia.com
What it says
The assigned deep dose equivalent must be for the part of the body receiving the highest exposure.
This quote was matched word for word to the source page.

Why this matters, in ORSIF's FAQs

Aprons and alternatives

Protective apparel, how it is checked, and whether other systems are addressed.

Must staff in Alaska fluoroscopy rooms wear protective aprons or other apparel?

Yes

Yes. Anyone other than the patient likely to be exposed to 5 mR per hour or more must wear an apron and gloves of at least 0.25 mm lead equivalent in the fluoroscopy room. Drapes or panels of at least 0.25 mm lead equivalent are also required between patient and fluoroscopist. No thyroid shield or eyewear requirement was found.

Rule text and source
Rule
7 AAC 18.440(l), (i) (eff. 4/9/2009)
Source
law.cornell.edu
What it says
Each person, other than the patient, whose body is likely to be exposed to five milliroentgens per hour or more must wear a protective apron and gloves of at least 0.25 millimeters lead equivalent material in the fluoroscopy room.
This quote was matched word for word to the source page.

Why this matters, in ORSIF's FAQs

Do the rules in Alaska address engineered protection systems used in place of aprons?

Not foundLess certain

Engineered radiation protection systems are not addressed by name. Under 18.440(i), mounted drapes, hinged shields and sliding panels may not replace an apron. A waiver section (7 AAC 18.405) exists but was not read.

Rule text and source
Rule
7 AAC 18.440(i); 7 AAC 18.405 (title only)
Source
law.cornell.edu
What it says
Protective drapes, hinges, or sliding panels, of at least 0.25 millimeters lead equivalent material, must be provided between the patient and fluoroscopist... These devices may not substitute for the wearing of a protective apron.
This quote was matched word for word to the source page.

About ALARA+

Who operates the equipment

Who may use fluoroscopy equipment and what training the rules ask for.

Who may operate fluoroscopy equipment in Alaska?

In partLess certain

No profession, license or permit was found in the rules read. The registrant must ensure equipment is operated only by an adequately instructed, competent individual. No supervision requirement was found. A 2022 vendor blog says Alaska has no X-ray operator training legislation; the statutes were not verified.

Rule text and source
Rule
7 AAC 18.410(b) (eff. 4/9/2009)
Source
law.cornell.edu
What it says
A registrant of medical or veterinary x-ray equipment shall assure that all x-ray equipment under the registrant's control is operated only by an individual who is adequately instructed in safe operating procedures and competent in safe use of the equipment.
This quote was matched word for word to the source page.

Why this matters, in ORSIF's FAQs

What training or credentials does Alaska require to operate fluoroscopy equipment?

Yes

The registrant must keep a training record for each operator. The minimum curriculum covers radiation safety and ALARA, personnel monitoring, equipment operation and dose control, quality control, and facility procedures. Fluoroscopy operators need at least 10 hours of instruction. No physician pathway, American Registry of Radiologic Technologists (ARRT) requirement, state permit or continuing education requirement was found.

Rule text and source
Rule
7 AAC 18.420(a)-(c) (eff. 4/9/2009)
Source
law.cornell.edu
What it says
An operator of medical fluoroscopic equipment must receive a minimum of 10 hours of instruction in the safe operation of the fluoroscope.
This quote was matched word for word to the source page.

Why this matters, in ORSIF's FAQs

Changes on the way, and recent ones

Rules, bills and guidance the research found from 2025 and 2026, with the ones that touch fluoroscopy teams first.

  • Proposed ruleNew chapter 7 AAC 19 for dental radiological equipment; fee changes (Online Public Notice 212715)

    Comments closed November 16, 2023; Cornell now lists Chapter 19, suggesting adoption

    Following SB 173, the proposal adopts Conference of Radiation Control Program Directors (CRCPD) Suggested State Regulations Part F (April 2015) by reference for dental equipment. It also changes the fee calculation for non-dental X-ray devices (7 AAC 80.030).

    Source: aws.state.ak.us

  • Proposed ruleProposed changes to 7 AAC 18 (lower dose limits)

    Appears not adopted; current text still shows 5 rem and 15 rem

    A DHSS draft from circa 2013 would have lowered the whole-body limit to 2 rem and the lens limit to 5 rem.

    Source: aws.state.ak.us

Also worth knowing

  • Fluoroscope equipment rules (18.440) require a five-minute cumulative timer, a continuous audible alarm in high-level control mode, a dead-man exposure switch, a Bucky-slot cover of at least 0.25 mm lead equivalent, and a record of patient dose (or exposure factors) for each procedure.
  • Doses from all employers count together (18.210(c)). Monitoring records must be kept for at least three years after a worker leaves (18.280(b)).
  • Cornell still lists DEC chapter 18 AAC 85 (Radiation Protection). A Cornell note says the 7 AAC 18 rules took effect in 2009 along with repeal of DEC rules on the same subject. Whether 18 AAC 85 still applies to anything relevant is unclear.
  • The program's dental inspection checklist asks facilities to describe how ALARA is practiced. That checklist relies on CRCPD Part F, which applies to dental, not medical, equipment.
  • No 2025 or 2026 proposals, bills or newly adopted rules affecting medical fluoroscopy were found, but the search was not complete.

What ORSIF could not confirm

Questions the research could not settle. If you know the answer, tell us.

  • Check every quote against the official Alaska Administrative Code at akleg.gov, and confirm nothing in 7 AAC 18 has changed since the 2025 Justia versions.
  • Read 7 AAC 18.255 for the declared pregnant worker limit, and the full 7 AAC 18.240 for the fetal dose limit in uncontrolled areas.
  • Read the unread sections: 7 AAC 18.405 (waivers; a possible route to approve apron alternatives), 18.270 (surveys), 18.300 (instruction of personnel) and 18.990 (definitions). These may affect the answers on dosimeters, apron inspection and apron alternatives.
  • Confirm in the statutes (AS 08; AS 18.60.475-.525) that Alaska has no radiologic technologist license or fluoroscopy permit. Check what the AS 18.60.525 exception in 7 AAC 18.105(a) covers, and read the Legislative Research Services report on X-ray operator standards (30th Legislature).
  • Determine the status of the circa-2013 proposal to lower the whole-body limit to 2 rem and the lens limit to 5 rem; the 2015 Register 214 amendments may show which parts were adopted.
  • Check the Online Public Notices system and the program (doh.radiation.control@alaska.gov, 907-371-1000) for 2025-2026 rule work on medical X-ray, and check whether 18 AAC 85 (DEC) is still in effect.

Double-check before you rely on this page

ORSIF compiled this page from the sources linked on it and did its best. It is not comprehensive and it is not always accurate. Rules change, a rule may have been missed or misread, and the sources themselves can be out of date. Treat this page as a starting point.

Where to check

  • The rule itself. Open "Rule text and source" under any answer, or read the sources listed below.
  • The state. Radiological Health Program. Program contacts in the CRCPD directory.
  • Your radiation safety officer. They know the federal, accreditation and facility requirements that apply at your site, which the state rules do not cover.

What may be missing here

  • 4 of 14 answers on this page are "Not found" or "Unclear". That means ORSIF did not find the answer in the rules it reviewed, not that no rule exists.
  • 7 answers carry a flag or a lower confidence ("Check the source" or "Less certain").
  • The research for this state ended before every question was fully answered, so some answers may be incomplete.
  • Researched November 3, 2026. Rules may have changed since.
  • Only the state's rules for X-ray and fluoroscopy equipment were reviewed.

Found something wrong or missing? Tell ORSIF.

Sources and terms

Terms used on this page
ALARA
As low as reasonably achievable: keep radiation doses as low as is reasonable, below the legal limits.
Agreement State
A state that has signed an agreement with the U.S. Nuclear Regulatory Commission to regulate certain radioactive materials itself.
mSv
Millisievert, a unit of radiation dose. Ten mSv equal one rem, the older unit some rules still use.
Dosimeter
A badge or device worn to measure a worker's radiation dose.
Lead equivalent
The thickness of lead (mm Pb) that would give the same protection as the material described.
NRC
U.S. Nuclear Regulatory Commission
ARRT
American Registry of Radiologic Technologists
CRCPD
Conference of Radiation Control Program Directors
All 19 sources the research read

Found something out of date or wrong? Tell ORSIF. Corrections are checked against the rule text and the page is updated.