Research summary, guideline or consensus statement, 2010

SIR-CIRSE Joint Guideline on Occupational Radiation Protection

Miller DL, Vañó E, Bartal G, Balter S, Dixon R, Padovani R, et al.. Cardiovascular and interventional radiology. 2010

Read the original paper

In plain language

This Society of Interventional Radiology (SIR)-Cardiovascular and Interventional Radiological Society of Europe (CIRSE) guideline advises the fluoroscopy team to combine aprons, thyroid shields, ceiling and table shields, and eye protection, wear two dosimeters, and use patient dose-reduction techniques, noting that properly protected interventional radiologists typically receive 2-4 mSv per year.

At a glance

Type of study
Guideline or consensus statement
Who was studied
Interventional radiologists, Nurses, Radiologic technologists, Anesthesia staff, Mixed lab staff
Summary based on
the full text

What it found

  • Eye protection (ceiling-suspended shields or leaded eyewear) is strongly advised at all times because the cataract threshold may be very low or absent and can be reached within several years.

    Key figures: threshold possibly <0.1 Gy

  • Disposable bismuth or tungsten-antimony patient drapes substantially reduce operator dose.

    Key figures: 12-fold eyes, 26-fold thyroid, 29-fold hands

  • Leaded glasses reduce frontal eye dose by about 8-10 times, but only 2-3 times when side exposure is included.

    Key figures: factor 8-10 frontal; 2-3 with side exposure

  • A busy interventional radiologist using appropriate precautions typically receives an effective dose of 2-4 mSv/year.

    Key figures: 2-4 mSv/year; unlikely >10 mSv/year

  • 0.5-mm lead-equivalent composite and lead-free aprons vary in protection; International Commission on Radiological Protection (ICRP) recommends two dosimeters (under and over apron) and annual fluoroscopic apron inspection.

    Key figures: 0.6% to 6.8% transmission

Limits to keep in mind

Expert consensus guideline based on a literature review without formal grading of evidence; cited dose figures come from other studies and mostly concern physicians rather than nurses, technologists or other staff.

How this summary was made

ORSIF prepared this summary with AI assistance from the full text of the paper. A program then checked that the passage supporting each finding appears word for word in that source. A person has not yet reviewed each summary, so read the original before relying on a number. ORSIF summarizes research and does not endorse any product or company. Report a problem with this summary.

Prepared November 3, 2026.