In plain language
In phantom tests and 177 cath lab procedures, adding a disposable vertical shield to standard shielding was associated with about 42% lower interventional cardiologist wrist dose. Wrist dosimetry only; not randomized and no health outcomes measured.
At a glance
- Type of study
- Original research, Prospective cohort study
- Who was studied
- Interventional cardiologists, Phantom or simulation (no people)
- Size
- n = 177 procedures
- Topics
- Room shielding and protective devicesDosimetry and monitoringDose reduction techniquesPatient dose
- Summary based on
- the full text
What it found
In clinical procedures with horizontal radiation absorbing pads, adding the vertical radiation shield lowered operator wrist dose by 41.9% with no difference in air kerma, and the difference persisted after multivariate adjustment.
Key figures: 41.9% reduction (15.2 ± 13.4 vs. 26.2 ± 31.4 μSv, p = .001); adjusted p < .001
In phantom testing, VRS added to horizontal pads reduced mannequin dose rate by 59% in one lab and dose by 100% in a second lab.
Key figures: 59% (p < .005); 100% (p = .016)
Combined HRAP and VRS reduced mannequin dose rate by 80% compared with no shielding in phantom model A.
Key figures: 80% (p < .005)
Left anterior oblique (LAO)-caudal angulation produced the highest operator and patient exposure, two to four times other angles.
Key figures: 2-4 times more radiation, p < .005
8-inch magnification increased air kerma and operator dose rate versus 10-inch; moving 40 cm from the access site lowered operator dose rate by 30%.
Key figures: +86.5% mGray, +12.2% μSv/min; -30% at 40 cm, -47% at 120 cm
Limits to keep in mind
Clinical comparison was non-randomized and unblinded with imbalanced groups (magnification, sex, fluoroscopy time), from one institution with four cardiologists; only wrist dose measured, not head or eye, and VRS tested at a single angle and magnification in phantoms.
Related research summaries
- Determinants of operator and patient radiation exposure during cardiac catheterization: Insights from the RadiCure (RADIation reduction during cardiac catheterization using real-timE monitoring) trial (2015, Randomized trial)
- Obesity as a determinant of radiation burden in the cardiac catheterization laboratory (2026, Narrative review)
- Radiation protection practices in diagnostic and interventional cardiology: Position paper from Portuguese Association of Interventional Cardiology (2025, Guideline or consensus statement)
- Radiation Safety in Chronic Total Occlusion Percutaneous Coronary Intervention (2025, Narrative review)
- Radiation Exposure, Training, and Safety in Cardiology (2024, Narrative review)
How this summary was made
ORSIF prepared this summary with AI assistance from the full text of the paper. A program then checked that the passage supporting each finding appears word for word in that source. A person has not yet reviewed each summary, so read the original before relying on a number. ORSIF summarizes research and does not endorse any product or company. Report a problem with this summary.
Prepared November 3, 2026.