Recording interventionalists' occupational dose by clinical procedure: a novel approach to radiation safety

Sanchez RM, Vaño Carruana E, Fernández Soto JM. CVIR endovascular. 2026

Read the original paper

In plain language

In one Spanish hospital, linking electronic dosimeters to each procedure showed interventional cardiologists and radiologists where their dose came from, such as spinal injections and biliary drainages, and flagged operators for feedback on shielding use. Benefit on dose was anecdotal.

At a glance

Type of study
Original research, Dosimetry measurement study
Who was studied
Interventional cardiologists, Interventional radiologists
Size
n = 3,373 procedures
Where
Spain
Summary based on
the full text

What it found

  • Median operator dose per procedure was 15.3 µSv in cardiology and 17.2 µSv in interventional radiology; median cumulative over-apron dose over 9 months was 2.50 and 3.28 mSv.

    Key figures: Cardiology 2.50 (1.0) mSv, 15.3 (4.1) µSv/procedure; interventional radiology (IR) 3.28 (1.2) mSv, 17.2 (10) µSv/procedure

  • Median operator-to-patient dose ratio was 0.2 µSv/Gy·cm2 in cardiology and 0.3 in radiology, lower than values reported by other studies.

    Key figures: 0.2 and 0.3 µSv/Gy·cm2

  • For one neuroradiologist, simple spinal infiltrations produced most of the operator's dose despite little patient dose, showing where protection could be improved.

    Key figures: 75% of operator dose vs 10% of patient kerma-area product (KAP)

  • Over 2024, 6 of 21 interventionalists triggered yellow alerts and none triggered red; two had annual over-apron doses of 8.2 and 8.4 mSv, about double colleagues' average.

    Key figures: 6 of 21 yellow; 8.2 and 8.4 mSv

  • Biliary drainages accounted for 17% of these two operators' dose but 2.5% of KAP, with dose/KAP ratios of 3-4 µSv/Gy·cm2; after feedback, one of the two considerably reduced most metrics in 2025.

    Key figures: 17% PDE vs 2.5% KAP; 3-4 µSv/Gy·cm2

Limits to keep in mind

Single-centre descriptive report of a prototype with 19-21 operators, no control group and anecdotal before/after evidence. Electronic dosimeters have pulsed-beam limitations, up to 5% data loss, and over-apron Hp(10) was used as a surrogate for eye lens dose.

How this summary was made

ORSIF prepared this summary with AI assistance from the full text of the paper. A program then checked that the passage supporting each finding appears word for word in that source. A person has not yet reviewed each summary, so read the original before relying on a number. ORSIF summarizes research and does not endorse any product or company. Report a problem with this summary.