Research summary, systematic review or meta-analysis, 2026

Operator radiation burden and periprocedural outcomes in robotic-assisted versus manual percutaneous coronary intervention: a meta-analysis

Xin Y, Wang L, Han F, Chang Q, Li C, Ding C, et al.. BMC surgery. 2026

Read the original paper

In plain language

Pooling 11 mostly observational studies, robotic-assisted percutaneous coronary intervention (PCI) was linked to lower primary operator radiation dose and lower patient dose metrics, with longer procedures and similar technical success. Patient outcome benefits are uncertain. Doses for other cath lab staff were not reported.

At a glance

Type of study
Systematic review or meta-analysis, Meta-analysis
Who was studied
Interventional cardiologists, Patients
Size
n = 11 studies (2,950 patients)
Summary based on
the full text

What it found

  • Operator radiation exposure was lower with robotic-assisted PCI than manual PCI, though heterogeneity was high.

    Key figures: MD = -29.97 µSv, 95% confidence interval (CI): -38.08 to -21.87, P < 0.00001; I² = 95%; 5 studies, 919 patients

  • Patient air kerma and dose-area product were lower with robotic PCI.

    Key figures: Air kerma MD = -222.48 mGy (95% CI -284.15 to -160.82); dose-area product (DAP) MD = -78.02 dGy·cm² (95% CI -120.45 to -35.59)

  • Fluoroscopy time was shorter with robotic PCI.

    Key figures: MD = -6.14 min, 95% CI: -9.41 to -2.88, P = 0.0002

  • Robotic PCI procedures took longer than manual PCI.

    Key figures: MD = 10.53 min, 95% CI: 0.55-20.51; P = 0.04

  • Technical success was similar; major adverse cardiac events (MACE) and 12-month mortality were lower with robotic PCI but considered hypothesis-generating.

    Key figures: Technical success odds ratio (OR) = 1.12 (0.68-1.85); MACE OR = 0.48 (0.24-0.93); mortality OR = 0.50 (0.27-0.94)

Limits to keep in mind

Ten of 11 studies were observational with limited sample sizes and high heterogeneity for operator dose, fluoroscopy time and duration; only five studies reported operator dose and lesion complexity was inconsistently reported.

How this summary was made

ORSIF prepared this summary with AI assistance from the full text of the paper. A program then checked that the passage supporting each finding appears word for word in that source. A person has not yet reviewed each summary, so read the original before relying on a number. ORSIF summarizes research and does not endorse any product or company. Report a problem with this summary.