Research summary, original research, 2018

Operator Dose Reduction during Transjugular Liver Biopsy Using a Radiation-Attenuating Drape: A Prospective, Randomized Study

Kohlbrenner R, Lehrman ED, Taylor AG, Kohi MP, Fidelman N, Kumar V, et al.. Journal of vascular and interventional radiology : JVIR. 2018

Read the original paper

In plain language

In a randomized trial of 62 transjugular liver biopsies, disposable radiation-absorbing drapes lowered the operator's measured chest dose by about half, without changing patient dose indicators. This is a dosimetry result, not a health outcome.

At a glance

Type of study
Original research, Randomized trial
Who was studied
Interventional radiologists
Size
n = 62 procedures
Summary based on
the abstract

What it found

  • Radiation-absorbing drapes cut mean operator chest dose by 56% compared with standard draping.

    Key figures: 37 μSv ± 35 vs 84 μSv ± 58; 56% reduction

  • Patient radiation measures did not differ between groups, suggesting procedures were comparable.

Limits to keep in mind

Single-institution randomized trial of 62 procedures limited to transjugular liver biopsy, measuring only the operator's chest dosimeter; no p-value for the dose difference is reported in the abstract, and other staff and body sites were not assessed.

How this summary was made

ORSIF prepared this summary with AI assistance from the abstract of the paper. A program then checked that the passage supporting each finding appears word for word in that source. A person has not yet reviewed each summary, so read the original before relying on a number. ORSIF summarizes research and does not endorse any product or company. Report a problem with this summary.

Prepared November 3, 2026.