Research summary, original research, 2016

EFFICACY OF AN ADDITIONAL MOBILE LEADED SHIELD ON OPERATOR RADIATION DURING PERCUTANEOUS

Luçon A, Ghrairi A, Lecoq G, Idali M, Huret B, Richard P, et al.. Radiation protection dosimetry. 2016

Read the original paper

In plain language

In one cath lab, adding a mobile leaded shield during radial diagnostic coronary angiography cut two cardiologists' measured eye dose by about a third to 43% and hand dose by 61-83%, without lengthening procedures. The study was very small.

At a glance

Type of study
Original research, Dosimetry measurement study
Who was studied
Interventional cardiologists
Size
n = 2 operators
Summary based on
the abstract

What it found

  • Adding a mobile leaded shield reduced operator eye dose in both operators.

    Key figures: 43% and 33% reduction (Operators 1 and 2)

  • The mobile shield reduced operator hand dose substantially.

    Key figures: 61% and 83% reduction (Operators 1 and 2)

  • Chest dose under the apron was below the minimum measurable dose for both operators.

    Key figures: <0.01 mSv

  • Using the shield did not change procedure duration, fluoroscopy time or contrast dose.

  • Nearly all procedures were done by radial access.

    Key figures: 99%

Limits to keep in mind

Only two operators in a single cath lab over two months, with a before-after (non-randomised) design; number of procedures not reported in the abstract and results are dosimeter readings, not health outcomes.

How this summary was made

ORSIF prepared this summary with AI assistance from the abstract of the paper. A program then checked that the passage supporting each finding appears word for word in that source. A person has not yet reviewed each summary, so read the original before relying on a number. ORSIF summarizes research and does not endorse any product or company. Report a problem with this summary.