Research summary, original research, 2013

A novel complete radiation protection system eliminates physician radiation exposure and leaded aprons

Fattal P, Goldstein JA. Indian Heart Journal. 2013

Read the original paper

In plain language

In a small nonrandomized series of 19 catheterization cases, operators working inside a complete shielding enclosure had no measurable significant radiation exposure, while those using standard shielding were exposed in all body regions. Health outcomes and other team members were not studied.

At a glance

Type of study
Original research, Dosimetry measurement study
Who was studied
Interventional cardiologists
Size
n = 19 procedures
Summary based on
the abstract

What it found

  • Operators using the Trinity system had no measurable significant radiation exposure in any body region or for the whole case.

  • Operators using standard shielding alone were exposed in all body regions, and total per-case exposure differed significantly from the Trinity system.

    Key figures: P < 0.0001

  • The system combines fixed shields, radiation drapes and flexible radiation-resistant materials while keeping full patient contact and equipment control, with the aim of removing the need for lead aprons.

Limits to keep in mind

Small, nonrandomized first-in-man series of 19 cases with exposure to operators only. Dose values, other staff exposure and ergonomic outcomes are not reported in the abstract, and the claim of eliminating aprons was not tested directly.

How this summary was made

ORSIF prepared this summary with AI assistance from the abstract of the paper. A program then checked that the passage supporting each finding appears word for word in that source. A person has not yet reviewed each summary, so read the original before relying on a number. ORSIF summarizes research and does not endorse any product or company. Report a problem with this summary.

Prepared November 3, 2026.