A dosimetry innovation using hybrid pixel detectors in interventional theatres

Genetay T, Otto T, Bochud F, Carbonez P, Damet J. Scientific reports. 2026

Read the original paper

In plain language

In laboratory reference X-ray beams, a CERN-derived pixel detector measured current and newly proposed staff dose quantities within 5% up to 70 kVp and about 20% off at 120 kVp. It has not yet been tested in clinical fluoroscopy rooms.

At a glance

Type of study
Original research, Dosimetry measurement study
Size
n = 8 beam qualities
Where
Switzerland
Summary based on
the full text

What it found

  • The Timepix4 hybrid pixel detector measured both current (H*(10)) and newly proposed International Commission on Radiation Units and Measurements (ICRU) 95 (H*) ambient dose quantities within 5% of a reference ionisation chamber up to 70 kVp.

    Key figures: within 5% up to 70 kVp

  • Accuracy fell at higher tube voltages, with deviations of about 20% at 120 kVp, attributed to fluorescence from bonding materials and low silicon sensor efficiency.

    Key figures: approximately 20% at 120 kVp

  • Repeat measurements were highly stable for a given beam quality.

    Key figures: standard deviation not higher than 0.5%

  • All measured values fell within IEC 60846-1 relative response limits, so the authors consider the detector already compliant with existing standards.

  • The authors note that conventional radiation protection detectors overestimate staff exposure by about a factor of two at diagnostic X-ray energies, motivating spectral dosimetry.

    Key figures: approximately a factor of two

Limits to keep in mind

Tested only in a calibration laboratory with continuous, unscattered standard IEC 61267 RQR/RQA beams at a non-reference 2.5 m distance, not in pulsed scattered fields in clinical rooms or worn by staff. Accuracy degrades above 70 kVp and the silicon sensor is unsuitable above 150 keV.

How this summary was made

ORSIF prepared this summary with AI assistance from the full text of the paper. A program then checked that the passage supporting each finding appears word for word in that source. A person has not yet reviewed each summary, so read the original before relying on a number. ORSIF summarizes research and does not endorse any product or company. Report a problem with this summary.