Technical image quality assessment as a tool for clinical audit in interventional radiology: first results of a National MoH Project

Zucca S, Serra G, Cavallari M, Postiglione F, Mazzocchi S, Pini S, et al.. Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB). 2026

Read the original paper

In plain language

In a phantom-based audit of 34 Italian interventional X-ray systems, dose and image quality varied widely, even between identical models. Systems with advanced noise reduction lowered dose while keeping image quality. Staff dose was not measured.

At a glance

Type of study
Original research, Phantom or simulation study
Who was studied
Phantom or simulation (no people)
Size
n = 34 X-ray systems (26 fixed angiographic, 8 mobile C-arm)
Where
Italy
Summary based on
the abstract

What it found

  • Image quality and patient entrance dose varied widely across systems and protocols, including among identical models.

  • Higher dose per frame generally improved image quality metrics, with exceptions such as digital subtraction angiography (DSA) cine mode.

  • Systems with advanced noise-reduction algorithms reduced dose while preserving image quality.

  • Limiting spatial resolution assessment showed high inter-observer agreement.

    Key figures: weighted Fleiss Kappa = 0.82

  • A standardized phantom-based audit across six medical physics centers was feasible, yielding 133 valid TCDD and 129 LSR datasets from 136 dosimetric sessions.

    Key figures: 136 sessions; 133 TCDD; 129 LSR

Limits to keep in mind

Phantom-based measurements of patient entrance dose and image quality only; no staff dose or clinical image quality was measured. Pilot audit limited to 34 systems in Italy.

How this summary was made

ORSIF prepared this summary with AI assistance from the abstract of the paper. A program then checked that the passage supporting each finding appears word for word in that source. A person has not yet reviewed each summary, so read the original before relying on a number. ORSIF summarizes research and does not endorse any product or company. Report a problem with this summary.