Borota L, Patz A. SPOT REGION OF INTEREST IMAGING: A NOVEL FUNCTIONALITY AIMED AT X-RAY DOSE REDUCTION IN NEUROINTERVENTIONAL PROCEDURES. Radiat Prot Dosimetry. 2020 Jun 24;188(3):322-331.

The aim of this study was to describe a new functionality aimed at X-ray dose reduction, referred to as spot region of interest (Spot ROI) and to compare it with existing dose-saving functionalities, spot fluoroscopy (Spot F), and conventional collimation (CC). The results for all FOVs were the following: for the fluoroscopy, all measured parameters for Spot ROI and Spot F were lower than corresponding values for CC. For DSA and DSA plus fluoroscopy, all measured parameters for Spot ROI were lower than corresponding parameters for Spot F and CC.